FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT II 2-PIECE SYRINGE

MDR report key: 16653046 · Received March 30, 2023

Report

Report Number
3002682307-2023-00088
Event Type
Malfunction
Date Received
March 30, 2023
Date of Event
March 13, 2023
Report Date
June 5, 2023
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2211257. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY (20) RETAINED SAMPLES WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR REVIEW; HOWEVER, NONE OF THE RETAINED SAMPLES DISPLAYED ANY SIGNS OF LEAKAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT II 2-PIECE SYRINGE EXPERIENCED (SHORT DESCRIPTION OF EVENT) THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: DURING THE WITHDRAWAL OF THE TREATMENT WITH THE SYRINGE, THE LATTER PASSED BETWEEN THE JOINT OF THE SYRINGE AND THE PLUNGER AND THE PRODUCT FLOWED OUT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT II 2-PIECE SYRINGE EXPERIENCED (SHORT DESCRIPTION OF EVENT) THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: DURING THE WITHDRAWAL OF THE TREATMENT WITH THE SYRINGE, THE LATTER PASSED BETWEEN THE JOINT OF THE SYRINGE AND THE PLUNGER AND THE PRODUCT FLOWED OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
580097 BD DISCARDIT II 2-PIECE SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2211257

Patients

Seq Age Sex Outcome Treatment
1 Unknown