FDA Adverse Event Malfunction Summary report: N

STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL

MDR report key: 1664630 · Received April 22, 2010

Report

Report Number
3005099803-2010-01830
Event Type
Malfunction
Date Received
April 22, 2010
Date of Event
February 19, 2010
Report Date
April 12, 2010
Manufacturer
ACCELLENT ENDOSCOPY JUAREZ
Product Code
FGO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT AGE IS REPORTED TO BE OVER 18 YEARS OLD. THE LOT NUMBER IS NOT KNOWN PER THE COMPLAINANT; THEREFORE, THE DEVICE MANUFACTURE AND EXPIRATION DATES CANNOT BE DETERMINED. (B)(6). EVALUATION OF THE RETURNED DEVICE INDICATES THE BLUE SHEATH WAS KINKED AND FISH HOOKED IN SEVERAL PLACES ALONG THE DISTAL END OF THE DEVICE, AND TEARING WAS OBSERVED ALONG THE WHITE HEAT SHRINK. THE DISTAL STOP WAS JAMMED INSIDE THE BLUE SHEATH, AND THE DEVICE WAS NOT ABLE TO BE OPENED. ADDITIONALLY, AN ACCORDION DEFECT WAS OBSERVED AT THE DISTAL TIP REVEALING A PORTION OF THE CORE WIRE. NO ANATOMICAL OR PROCEDURAL FACTORS WERE NOTED WHICH MIGHT HAVE CAUSED OR CONTRIBUTED TO THIS COMPLAINT. THE MOST PROBABLE ROOT CAUSE FOR THIS COMPLAINT IS UNDETERMINABLE.

Description of Event or Problem · 1

NOTE: THIS EVENT, AS REPORTED, DID NOT REFLECT AN MDR REPORTABLE SCENARIO; HOWEVER, EVALUATION OF THE RETURNED DEVICE REVEALED AN MDR-REPORTABLE MALFUNCTION. THIS MANUFACTURER REPORT IS BEING SUBMITTED BASED ON THE EVALUATION RESULTS. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON FEBRUARY 24, 2010 THAT A STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL WAS USED IN A "REMOVAL OF THE BILE DUCT STONES" PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, THE DEVICE WAS UNABLE TO OPEN AND CLOSE INSIDE THE PATIENT'S URETER. NO STONE WAS IN THE DEVICE WHEN THE PROBLEM OCCURRED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "GOOD."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL DISLODGER, STONE, FLEXIBLE FGO ACCELLENT ENDOSCOPY JUAREZ M0063903200

Patients

Seq Age Sex Outcome Treatment
1