I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2023-00027
- Event Type
- Malfunction
- Date Received
- March 30, 2023
- Date of Event
- March 19, 2023
- Report Date
- April 19, 2023
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749001917
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). ADDITIONAL PRODUCT TO BE INVESTIGATED. DESCRIPTION: CG8+. LOT#: W22291. PRODUCT #: 600-9001-25. PMA/510K#: K940918. UDI: (B)(4). MFG DATE: 18-OCT-2022. EXPIRY DATE: 23-JUN-2023. APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4) THE INVESTIGATION WAS COMPLETED ON 18-MAR-2023. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AK, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H22316 OR CG8+ CARTRIDGE LOT W22291.
ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED DISCREPANT POTASSIUM RESULTS ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. METHOD, DATE TESTED, RESULT, SAMPLE, CARTRIDGE, LOT#: I-STAT (B)(6) 2023 10:46 2.0 MEQ/L B CHEM8+ H22316; VS2 (B)(6) 2023 02:15 2.6 MMOL/L B CHEM8+ H22316; I-STAT (B)(6) 2023 20:39 3.0 MEQ/L C CG8+ W22291;VS2 (B)(6) 2023 21:21 4.6 MMOL/L C CG8+ W22291; I-STAT (B)(6) 2023 02:51 3.1 MEQ/L A CHEM8+ H22316; VS2 (B)(6) 2023 03:39 4.2 MMOL/L A CHEM8+ H22316. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 591591 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE, | JGS | ABBOTT POINT OF CARE | NA | H22316 | 10054749001917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |