FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 16646249 · Received March 30, 2023

Report

Report Number
2245578-2023-00027
Event Type
Malfunction
Date Received
March 30, 2023
Date of Event
March 19, 2023
Report Date
April 19, 2023
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749001917
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). ADDITIONAL PRODUCT TO BE INVESTIGATED. DESCRIPTION: CG8+. LOT#: W22291. PRODUCT #: 600-9001-25. PMA/510K#: K940918. UDI: (B)(4). MFG DATE: 18-OCT-2022. EXPIRY DATE: 23-JUN-2023. APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4) THE INVESTIGATION WAS COMPLETED ON 18-MAR-2023. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AK, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOT H22316 OR CG8+ CARTRIDGE LOT W22291.

Description of Event or Problem · 0

ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED DISCREPANT POTASSIUM RESULTS ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. METHOD, DATE TESTED, RESULT, SAMPLE, CARTRIDGE, LOT#: I-STAT (B)(6) 2023 10:46 2.0 MEQ/L B CHEM8+ H22316; VS2 (B)(6) 2023 02:15 2.6 MMOL/L B CHEM8+ H22316; I-STAT (B)(6) 2023 20:39 3.0 MEQ/L C CG8+ W22291;VS2 (B)(6) 2023 21:21 4.6 MMOL/L C CG8+ W22291; I-STAT (B)(6) 2023 02:51 3.1 MEQ/L A CHEM8+ H22316; VS2 (B)(6) 2023 03:39 4.2 MMOL/L A CHEM8+ H22316. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
591591 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE, JGS ABBOTT POINT OF CARE NA H22316 10054749001917

Patients

Seq Age Sex Outcome Treatment
1 Unknown