PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Report
- Report Number
- 2024168-2023-03248
- Event Type
- Injury
- Date Received
- March 30, 2023
- Date of Event
- March 6, 2023
- Report Date
- May 11, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS OBSERVED AS A LINK BREAK. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO DESIGN, MANUFACTURE OR LABELING OF THE DEVICE.D4 - LOT # UPDATED FROM 2101741 TO 2101742.
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PROSTYLE DEVICES WITH REPORTED ISSUES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.
IT WAS REPORTED THAT SUTURE PLACEMENT IN THE LEFT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A PROSTYLE DEVICE USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURE WAS NOT PRESENT WHEN THE PLUNGER WAS REMOVED WITH SIX PROSTYLE DEVICES (A CUFF MISS OCCURRED). THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 20F AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 177860 | PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 2101742 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |