FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 16645966 · Received March 30, 2023

Report

Report Number
2024168-2023-03249
Event Type
Injury
Date Received
March 30, 2023
Date of Event
March 6, 2023
Report Date
May 11, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS OBSERVED AS A LINK BREAK. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO DESIGN, MANUFACTURE OR LABELING OF THE DEVICE.D4 - LOT # UPDATED FROM 2101741 TO 2101742.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PROSTYLE DEVICES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Description of Event or Problem · 0

IT WAS REPORTED THAT SUTURE PLACEMENT IN THE LEFT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A PROSTYLE DEVICE USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURE WAS NOT PRESENT WHEN THE PLUNGER WAS REMOVED WITH SIX PROSTYLE DEVICES (A CUFF MISS OCCURRED). THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 20F AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
566476 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 2101742

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention