FDA Adverse Event Malfunction Summary report: N

PASSEO-18 5/120/130

MDR report key: 16645720 · Received March 30, 2023

Report

Report Number
1028232-2023-01601
Event Type
Malfunction
Date Received
March 30, 2023
Date of Event
March 6, 2023
Report Date
March 29, 2023
Manufacturer
BIOTRONIK AG, BUELACH, SWITZERLAND
Product Code
LIT
UDI-DI
07640119559223
PMA / PMN Number
K072765
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NEITHER THE COMPLAINT INSTRUMENT NOR THE ANGIOGRAPHIC MATERIAL WAS RETURNED. THEREFORE, NO TECHNICAL INVESTIGATION ON THE SUBJECT COULD BE PERFORMED. THE PRODUCT RELEASE DOCUMENTATION WAS REVIEWED TO ESTABLISH WHETHER A DEVIATION FROM THE MANUFACTURING PROCESS COULD BE THE CAUSE FOR THE REPORTED EVENT. REVIEW OF THE PRODUCT RELEASE DOCUMENTATION FOR THE PRODUCT DETAILED ABOVE VERIFIED THAT THE INSTRUMENT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS AND FULFILLED ALL THE REQUIREMENTS OF IN-PROCESS AND FINAL INSPECTION. BASED ON THE CONDUCTED INVESTIGATIONS, NO MATERIAL OR MANUFACTURING RELATED ROOT CAUSE COULD BE DETERMINED.

Description of Event or Problem · 0

THE PASSEO-18 BALLOON CATHETER WAS CHOSEN TO CONDUCT THE PRE-DILATATION OF A MILDLY CALCIFIED LESION WITH 90 PERCENT STENOSIS DEGREE IN THE MODERATELY TORTUOUS SFA, BUT THE LESION COULD NOT BE PASSED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
567278 PASSEO-18 5/120/130 PERIPHERAL DILATATION CATHETER LIT BIOTRONIK AG, BUELACH, SWITZERLAND 357511 05221181 07640119559223

Patients

Seq Age Sex Outcome Treatment
1 Unknown