X-SMART W/CONTRA ANGLE EUR
Report
- Report Number
- 8031010-2023-00774
- Event Type
- Malfunction
- Date Received
- March 30, 2023
- Report Date
- April 13, 2023
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- EKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
RECEIVED = 1X X-SMART CHARGER EUR A100700000200. 1X X-SMART CONTRA ANGLE 16:1 A100601600000 SN: (B)(6) -(B)(6). 1X X-SMART HAND PIECE A100500000100 SN: (B)(6). 1X X-SMART UNIT SN: (B)(6). X-SMART MOTOR ASSEMBLY. SAV VARIOUS MECHANICAL PROBLEM. THE INSIDE SCREW OF MOTOR IS BROKEN, THEREFORE THE MOTOR HEAD IS NO LONGER FIXED TO THE MOTOR BODY. X-SMART HEAD CAP. SAV BAD MAINTENANCE (USER). THERE IS SOME FOREIGN MATTER ON THE HEAD CAP. X-SMART BATTERY. SAV BATTERY LEVEL TOO LOW TO BE RECHARGED. THE BATTERY VOLTAGE IS TOO LOW. WITH THIS VOLTAGE: IT IS IMPOSSIBE TO CHARGE THE BATTERY AND IT IS IMPOSSIBLE TO TURN ON THE DEVICE. THE CAUSE OF THIS PROBLEM IS PROBABLY A TOO LONG STORAGE. X-SMART CARTRIDGE. SAV VARIOUS MECHANICAL PROBLEM I NOTE THAT THE CARTRIGDE IS BLOCKED.. X-SMART NECK. SAV VARIOUS MECHANICAL PROBLEM. FOREIGN MATTER. REPLACED 1X X HAND PIECE A100500000100 (B)(6). 1X X CHARGER EUR A100700000200. 1X X CONTRA ANGLE 16:1 A100601600000 (B)(6). 1X X UNIT (B)(6).
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT IT IS REPORTED THAT A X-SMART W/CONTRA ANGLE EUR AUTOREVERSE WAS DEFECTIVE. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173438 | X-SMART W/CONTRA ANGLE EUR | HANDPIECE, DIRECT DRIVE, AC-POWERED | EKX | MAILLEFER INSTRUMENTS HOLDING SARL | 08142243-08142433, 08232730, 0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |