FDA Adverse Event Malfunction Summary report: N

MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 16641919 · Received March 30, 2023

Report

Report Number
2032227-2023-178397
Event Type
Malfunction
Date Received
March 30, 2023
Date of Event
January 12, 2023
Report Date
March 30, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. H6- COMPONENT CODE WAS WRONG IN THE INITIAL REPORT. CORRECT H6- COMPONENT CODE IS BEING SUBMIT VIA THIS FOLLOW UP REPORT UNDER H6.

Additional Manufacturer Narrative · 0

(B)(4). PUMP PASSED THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, ACTIVE CURRENT TEST, SELF-TEST, AND DISPLACEMENT ACCURACY TEST BUT FAILED THE SLEEP CURRENT TEST, PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLY. THUMP WAS UTILIZED AND DOWNLOADED TRACE/HISTORY FILES PROPERLY. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WAS WITHIN SPEC RANGE. NO UNEXPECTED BATTERY ALARMS/ALERTS DURING TESTING OR IN THE PUMP HISTORY. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO PHYSICAL COMPONENT OR MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, HARNESS VIBRATOR ASSEMBLY, BATTERY TUBE, MOTOR, OR FORCE SENSOR. SWAPPING OUT TEST BOARDS CONFIRMS FAILED SLEEP CURRENT MEASUREMENT, PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLY. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: BROKEN BELT CLIP RAILS, PILLOWING KEYPAD OVERLAY, AND STAINED SERIAL NUMBER LABEL DAMAGE. PUMP PASSED ALL OTHER REQUIRED TESTING BUT FAILED THE SLEEP CURRENT TEST. UNEXPECTED BATTERY POWER LOSS WAS CONFIRMED DUE TO HIGH SLEEP CURRENT, PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLY. UNABLE TO VERIFY CUSTOMER ALLEGED FOR HIGH BG. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 780G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE NGP INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEY HAD HIGH BLOOD GLUCOSE VALUE WAS 390 MG/DL AND CURRENT BLOOD GLUCOSE VALUE WAS 180 MG/DL. CUSTOMER REPORTED THAT THE AUTO MODE WAS ACTIVE AT THE TIME OF INCIDENT. CUSTOMER HAD UNDER DELIVERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
442517 MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1886

Patients

Seq Age Sex Outcome Treatment
1 Unknown