FDA Adverse Event Injury Summary report: N

COOLSCULPTING® SYSTEM

MDR report key: 16640528 · Received March 29, 2023

Report

Report Number
3007215625-2023-00405
Event Type
Injury
Date Received
March 29, 2023
Date of Event
March 14, 2023
Report Date
August 15, 2023
Manufacturer
ALLERGAN PLEASANTON
Product Code
OOK
PMA / PMN Number
K160259
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION (PROVIDED FROM CEF RECEIVED AFTER THE INITIAL EMDR): A2 (DOB), A3 (GENDER), A4 (WEIGHT), A4 (WEIGHT UNITS), A6 (RACE), D1 (BRAND NAME), CONCOMITANT PRODUCT (TWO APPLICATORS), AND B5 (COMPLAINT DESCRIPTION).

Additional Manufacturer Narrative · 0

ACCORDING TO THE COOLSCULPTING USER MANUAL, UNDER RARE ADVERSE EVENTS, PARADOXICAL ADIPOSE HYPERPLASIA (PH/PAH) IS CHARACTERIZED BY A VISIBLY ENLARGED TISSUE VOLUME WITHIN THE TREATMENT AREA, WHICH MAY DEVELOP TWO TO FIVE MONTHS AFTER TREATMENT. SURGICAL INTERVENTION MAY BE REQUIRED. PAH IS NOT RELATED TO ANY COOLSCULPTING DEVICE FAILURE MODE BUT IT IS INCLUDED IN THE RISK MANAGEMENT FILES OF THE DEVICE BECAUSE IT IS A RISK THAT IS INHERENT TO THE USE CRYOLIPOLYSIS FOR LOCALIZED FAT REDUCTION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF AND WHEN ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 0

ALLERGAN AESTHETICS RECEIVED A REPORT OF A PATIENT TREATED WITH COOLSCULPTING TO THE INFRA-SCAPULAR AREA ON (B)(6) 2022 USING THE COOLADVANTAGE+ COOLCURVE AND DEVELOPED PARADOXICAL HYPERPLASIA (PAH/PH) THREE MONTHS AFTER TREATMENT. DIAGNOSED ON (B)(6) 2023 WITH PARADOXICAL ADIPOSE HYPERPLASIA (PAH) BY MEDICAL PROFESSIONAL.

Description of Event or Problem · 0

ALLERGAN AESTHETICS RECEIVED A REPORT THAT A PATIENT RECEIVED A COOLSCULPTING TREATMENT AND PRESENTED WITH PARADOXICAL HYPERPLASIA (PAH/PH).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571675 COOLSCULPTING® SYSTEM DERMAL COOLING PACK/VACUUM/MASSAGER OOK ALLERGAN PLEASANTON NI

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female Required Intervention 2014316003| 2014318006