FDA Adverse Event Malfunction Summary report: N

TRITON CANISTER

MDR report key: 16632337 · Received March 28, 2023

Report

Report Number
3011157718-2023-00017
Event Type
Malfunction
Date Received
March 28, 2023
Date of Event
November 7, 2019
Report Date
March 28, 2023
Manufacturer
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
Product Code
PBZ
PMA / PMN Number
K142801
Removal / Correction Number
Z-1186-2023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

PER THE CUSTOMER, DURING THE CASE THE TEAM FELT THE EBL WAS VERY LOW AND UPON REVIEW I SEE 18 SPONGES SCANNED WITH A EBL OF 3 FROM 3 SPONGES. IT APPEARS THAT TWO SPONGES WERE OVER SATURATED BUT THE OTHER 13 SPONGES GAVE NO READING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412 TRITON CANISTER IMAGE PROCESSING DEVICE FOR ESTIMATION OF EXTERNAL BLOOD LOSS PBZ STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP

Patients

Seq Age Sex Outcome Treatment
1 Unknown