PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2023-03142
- Event Type
- Injury
- Date Received
- March 28, 2023
- Date of Event
- March 11, 2023
- Report Date
- June 19, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED FOR EVALUATION. THE UNSPECIFIED FAILURE WAS OBSERVED AS SUTURE RETRIEVAL ISSUE [NEEDLE TO CUFF MISS AND SUTURE SEPARATION]. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. D4 LOT# UPDATED FROM 2082441 TO: 2072641. D4: EXPIRATION DATE. H4: MANUFACTURING DATE.
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PROGLIDE DEVICE WITH REPORTED ISSUE REFERENCED IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH TWO PROGLIDE DEVICES AFTER A PROCEDURE USING A SMALL BORE HOLE. REPORTEDLY, UNSPECIFIED FAILURES OCCURRED WITH BOTH DEVICES. A NEW PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10599 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 2072641 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Unknown | Required Intervention |