SYNCHRON LX® I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2010-00201
- Event Type
- Malfunction
- Date Received
- April 20, 2010
- Date of Event
- March 24, 2010
- Report Date
- April 20, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K023049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
CALIBRATION AND QC ARE RUN EVERY 8 HOURS. QC PRIOR TO AND AFTER THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT, BUT FOUND NO PROBLEMS. THE FSE PERFORMED A PRECISION RUN ON NA, WHICH PASSED THE SPECIFICATIONS. A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING A FALSE LOW SODIUM (NA) RESULT GENERATED BY THE SYNCHRON LXI 725 CLINICAL SYSTEM. THE ORIGINAL RESULT OF 117 MMOL/L WAS REPORTED OUTSIDE OF THE LAB AND WAS QUESTIONED BY THE PHYSICIAN. THE SAMPLE WAS RUN ON ANOTHER INSTRUMENT AND THE RESULT OBTAINED WAS 138 MMOL/L. THE ORIGINAL RESULT WAS AMENDED. IT IS UNKNOWN IF TREATMENT WAS INITIATED OR WITHHELD BASED UPON THE FALSE LOW RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LX® I 725 CLINICAL SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JGS | BECKMAN COULTER INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |