BD SYNAPSYS¿
Report
- Report Number
- 1119779-2023-00331
- Event Type
- Malfunction
- Date Received
- March 28, 2023
- Date of Event
- February 20, 2023
- Report Date
- May 10, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- GWN
- UDI-DI
- 00382904441500
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE CUSTOMER REPORTED A SYNAPSES ISSUE WHERE, WHEN THEY CLICK "MARK AS READ" AND A NEW CULTURE POPS UP, 5-10 SECONDS GO BY BEFORE THE NEW ACCESSION NUMBER AND PLATE DEMOGRAPHICS SHOW UP. THIS WAS RECORDED AGAINST SYNAPSES, MATERIAL NUMBER 444150, SERIAL NUMBER SL00005. THIS WAS DETERMINED TO BE A BUG IN SYNAPSES WHERE THERE IS A BUG IN SYNAPSES WHICH CAN RESULT IN AN ASYNCHRONOUS LOADING BETWEEN THE CULTURE IMAGES AND THE PATIENT AND SPECIMEN INFORMATION PANEL. A CORRECTIVE AND PREVENTATIVE ACTION WAS OPENED TO INVESTIGATE THE ROOT CAUSE OF THIS PROBLEM. IN THIS CASE, THE CUSTOMER REPORTED THE INCORRECT INFORMATION TO THEIR LIS. THIS IS A CONFIRMED FAILURE OF A BD PRODUCT, THE ROOT CAUSE IS A BUG IN SYNAPSES. SERVICE UPDATED THE CUSTOMER TO THE LATEST VERSION OF SYNAPSES AND PERFORMED A DATABASE PURGE TO FREE UP RESOURCES AND MINIMIZE THE CHANCE FOR THIS DELAY IN LOADING. BD WILL CONTINUE TO MONITOR FOR TRENDS RELATED TO THIS ISSUE. H3 OTHER TEXT : SEE H.10.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD SYNAPSES¿ A SLOWNESS OF ACCESSION NUMBERS POPULATING IS NEGATIVELY AFFECTING PATIENT RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS SYNAPSES ISSUE THAT AN ISSUE WHERE WHEN THEY CLICK MARK AS READ AND A NEW CULTURE POPS UP THE PLATES SHOW UP, THEN 5-10 SECONDS GO BY BEFORE THE NEW CID AND PATIENT DEMOS SHOW UP.
IT WAS REPORTED THAT BD SYNAPSYS¿ A SLOWNESS OF ACCESSION NUMBERS POPULATING IS NEGATIVELY AFFECTING PATIENT RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS SYNAPSYS ISSUE THAT AN ISSUE WHERE WHEN THEY CLICK MARK AS READ AND A NEW CULTURE POPS UP THE PLATES SHOW UP, THEN 5-10 SECONDS GO BY BEFORE THE NEW CID AND PATIENT DEMOS SHOW UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 13192 | BD SYNAPSYS¿ | NA | GWN | BECTON, DICKINSON & CO. (SPARKS) | 00382904441500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |