FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1662189
·
Received April 12, 2010
Report
- Report Number
- 1824206-2010-05507
- Event Type
- Malfunction
- Date Received
- April 12, 2010
- Date of Event
- March 18, 2010
- Report Date
- March 18, 2010
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECH FOUND THE PILOT INSIDE THE SOLENOID VALVE. HAD HYD OIL BEHIND IN IT CAUSED BY A BAD/WORN OUT O-RING. HE REPLACED THE PILOT AND THE O-RING. ACCOUNT SUPPLIED THE PARTS. BED IS WORKING FINE NOW.
Description of Event or Problem · 1
TECH ALLEGED THE BED HEAD FUNCTION WOULDN'T WORK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM RITTER | P1900E004960 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |