CAPSURE SP
Report
- Report Number
- 2649622-2010-01879
- Event Type
- Death
- Date Received
- April 20, 2010
- Date of Event
- May 25, 2009
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S12
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B) (4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. (B) (4) INNER INSULATION SEPARATION. PROXIMAL SEGMENT RETURNED AND ANALYZED.
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THE LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. FOLLOW UP WITH THE CLINIC REPORTED THERE WERE NO DEVICE ALLEGATIONS AND THE PATIENT DIED OF NATURAL CAUSES AND WAS ON HOSPICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4524 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Other | 4524 IMPLANTABLE PACING LEAD| 4965 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD |