Narratives

Additional Manufacturer Narrative · 0

UPDATED DATA: B4,G3,G6,H2,H10,H11. CORRECTED DATA: B5,B6,B7,D10,E3,H6(CLINICAL & IMPACT).

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4, G3, G6, H2, H4, H6, H10.

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: TEL - (B)(6). IT WAS BEING REPORTED THAT DURING A ROUTINE CHECK THE CS300 INTRA AORTIC BALLOON PUMP (IABP) HAD AN ISSUE REGARDING THE "HELIUM LEAK" . A GETINGE FIELD SERVICE ENGINEER (FSE) HAD COMMENTED THAT THE THERE WAS A FOLLOW-UP CALL FOR HELIUM LEAK ON THE SHELL SIDE. SHELL ATTACHMENT CHECK: RAS SEAL CHECK: NO FAULT LEAK CHECK IN DIAGNOSTIC MODE: NO LOSS OVER 5 MINUTES. MONITORING AGREEMENT BY THE BIOMEDICAL SERVICE OVER SEVERAL DAYS FBC 3395. THERE WAS NO INVOLVEMENT OF THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USER CHECK THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) BOTTLE LEAKED. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) BOTTLE LEAKED.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
739939 CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-3023-62 10607567108308

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown N/A.| UNKNOWN.