FDA Adverse Event Injury Summary report: N

IMP,TSV,MCOL MG,4.7MM,10M

MDR report key: 16615503 · Received March 24, 2023

Report

Report Number
0002023141-2023-00911
Event Type
Injury
Date Received
March 24, 2023
Date of Event
February 27, 2023
Report Date
October 8, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019775
PMA / PMN Number
K111889
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE RECEIVED ONE (1) TSVMWB10, (IMP,TSV,MCOL MG,4.7MM,10M) FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, BONE ON THREADS AND A FRACTURE HAS BEEN IDENTIFIED ON THE IMPLANT COLLAR. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1232608. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1232608 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW AS PER RM-00541-HAZ, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WERE MATERIAL SELECTION OF IMPLANT INADEQUATE (UNABLE TO WITHSTAND OCCLUSAL FORCES OR LONG TERM LOADING), CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING, MISSING OR CONFUSING INSTRUCTIONS FOR USE. REFER TO ATTACHED SUMMARY INVESTIGATION FOR INFECTION. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. SIMILAR COMPLAINTS FOR IMPLANT INFECTION HAVE BEEN PREVIOUSLY INVESTIGATED. REFER TO ATTACHED SUMMARY INVESTIGATION. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORD REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH DID NOT SUGGEST THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/ OR ANY FURTHER ESCALATIONS. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. D9: DEVICE AVAILABILITY. G3: DATE RECEIVED BY MANUFACTURER. G6: CHECKED "FOLLOW-UP". H2: CHECKED FOLLOW-UP TYPE. H3: CHANGED "NO" TO "YES". H6: ENTERED EVALUATION CODES. H10: ADDED MANUFACTURER NARRATIVE.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED. AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. ADDITIONAL PMA/510(K) NUMBER ¿ K101880. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION ON TOOTH #19 AND TENDERNESS AT THE REGION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
486581 IMP,TSV,MCOL MG,4.7MM,10M DENTAL IMPLANT DZE ZIMMER DENTAL TSVMWB10 1232608 00889024019775

Patients

Seq Age Sex Outcome Treatment
1 Prefer Not To Disclose Required Intervention