FDA Adverse Event
Malfunction
Summary report: N
AED PRO
MDR report key: 16611959
·
Received March 24, 2023
Report
- Report Number
- 1220908-2023-01018
- Event Type
- Malfunction
- Date Received
- March 24, 2023
- Report Date
- March 3, 2023
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- P160022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION FOR EVALUATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A DAMAGED PINS ON A CONNECTOR LOCATED ON THE FLEX CIRCUIT ASSEMBLY. THE FLEX CIRCUIT ASSEMBLY WAS REPLACED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 0
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 521312 | AED PRO | DEFIBRILLATOR | MKJ | ZOLL MEDICAL CORPORATION | AED PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |