FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 16610487 · Received March 24, 2023

Report

Report Number
1000113657-2023-00166
Event Type
Malfunction
Date Received
March 24, 2023
Date of Event
March 1, 2023
Report Date
April 26, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
0002129200783
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 26-APR-2023: H6: UPDATED FDA¿S TYPE, FINDINGS AND CONCLUSIONS CODES. H10: RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND DEFECT FOUND ON RETURNED TEST STRIPS: HIGH BLOOD READINGS. NO DEFECT FOUND ON RETURNED METER. ROOT CAUSE: RC-061: STORAGE OUTSIDE SPECIFICATIONS. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 20-MAR-2023 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH AND ERRATIC BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 161, 131, 127, 350 AND 261 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 91-108 MG/DL AND EXPECTED TWO HOUR POST MEAL FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 111-132 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 241 MG/DL AND 239 MG/DL USING TRUE METRIX METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 04/25/2024 AND TEST STRIPS WERE OPENED THREE WEEKS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 161 MG/DL, DATE: (B)(6), TIME: 09:37 PM, FASTING 2 HOUR AFTER EATING. RESULT 2: 131 MG/DL, DATE: (B)(6), TIME: 09:28 PM, FASTING 2 HOUR AFTER EATING. RESULT 3: 127 MG/DL, DATE: (B)(6), TIME: 06:07 PM, FASTING 2 HOUR AFTER EATING. RESULT 4: 350 MG/DL, DATE: (B)(6), TIME: 06:00 PM, FASTING 2 HOUR AFTER EATING. RESULT 5: 261 MG/DL, DATE: (B)(6), TIME: 06:05 AM, FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375025 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX MEDIMEDI 50CT MG/DL ZA4950S 0002129200783

Patients

Seq Age Sex Outcome Treatment
1 Unknown