FDA Adverse Event Injury Summary report: N

THE SPANNER TEMPORARY PROSTATIC STENT

MDR report key: 16603272 · Received March 23, 2023

Report

Report Number
1000305746-2023-00005
Event Type
Injury
Date Received
March 23, 2023
Date of Event
February 23, 2015
Report Date
March 23, 2023
Manufacturer
SRS MEDICAL
Product Code
NZC
PMA / PMN Number
P060010
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WHEN THIS EVENT WAS INITIALLY RECEIVED, OUR PROCEDURES FOR DETERMINING MDR DID NOT CHARACTERIZE THIS EVENT AS REPORTABLE IN ACCORDANCE WITH 21 CFR 803.3(W). WE HAVE RECENTLY REVISED OUR PROCEDURES TO BROADEN THE INTERPRETATION OF 21 CFR 803.3(W)(3) AND ARE RETROSPECTIVELY REPORTING PREVIOUS EVENTS DATING BACK TO 2013.

Description of Event or Problem · 0

A TEMPORARY PROSTATIC STENT WAS PLACED BY A UROLOGIST ON (B)(6) 2015. ON (B)(6) 2015, THE PATIENT REPORTED URINARY RETENTION. ON (B)(6) 2015, THE UROLOGIST REMOVED THE STENT AND NOTED THAT THE STENT'S BALLOON WAS ALREADY DEFLATED AND THE STENT WAS PARTIALLY EXPELLED INTO THE PENILE URETHRA. THE PATIENT REPORTED ACCIDENTALLY PULLING THE STENT'S REMOVAL TETHER AND IT IS BELIEVED THAT THE PATIENT DEFLATED THE BALLOON. THERE WAS NO REPORTED PATIENT INJURY BUT THE STENT DID NOT ALLEVIATE THE PATIENT'S RETENTION SYMPTOMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1444685 THE SPANNER TEMPORARY PROSTATIC STENT STENT, URETHRAL, PROSTATIC, SEMI-PERMANENT NZC SRS MEDICAL SPNR-9HA WP01337

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention