FDA Adverse Event Malfunction Summary report: N

PUREPOINT LASER

MDR report key: 16597142 · Received March 23, 2023

Report

Report Number
2028159-2023-00341
Event Type
Malfunction
Date Received
March 23, 2023
Date of Event
February 22, 2023
Report Date
May 12, 2023
Manufacturer
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Product Code
HQF
UDI-DI
00380657505975
PMA / PMN Number
K062624
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE COMPANY REPRESENTATIVE WAS ABLE TO CONFIRM AND REPLICATE THE REPORTED EVENT. THE RADIO FREQUENCY IDENTIFICATION (RFID) FRONT PANEL PRINTED CIRCUIT BOARD (PCB) WAS REPLACED TO ADDRESS THE ISSUE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A NON-CONFORMANCE BASED REVIEW OF THE SERIAL NUMBER WAS PERFORMED AND DID NOT REVEAL ANY POTENTIAL CONTRIBUTING FACTORS TO THE REPORTED COMPLAINT. A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. A REVIEW FOR COMPLAINTS REPORTED AGAINST THIS SERIAL NUMBER WAS PERFORMED. NO SIMILAR COMPLAINTS WERE REPORTED FOR THE PRODUCT SERIAL UNDER INVESTIGATION. THE ROOT CAUSE OF THE REPORTED EVENT IS ATTRIBUTED TO THE NONCONFORMING RFID PCB. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A CUSTOMER REPORTED THAT AN OPHTHALMIC CONSOLE EXHIBITED LOW LASER POWER. THE DETAILS ABOUT PROCEDURE AND PATIENT CONTACT WAS NOT REPORTED. ADDITIONAL INFORMATION HAD BEEN REQUESTED BUT NONE RECEIVED TILL DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3297 PUREPOINT LASER LASER, OPHTHALMIC HQF ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER NA NA 00380657505975

Patients

Seq Age Sex Outcome Treatment
1 Unknown