PUREPOINT LASER
Report
- Report Number
- 2028159-2023-00341
- Event Type
- Malfunction
- Date Received
- March 23, 2023
- Date of Event
- February 22, 2023
- Report Date
- May 12, 2023
- Manufacturer
- ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
- Product Code
- HQF
- UDI-DI
- 00380657505975
- PMA / PMN Number
- K062624
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
THE COMPANY REPRESENTATIVE WAS ABLE TO CONFIRM AND REPLICATE THE REPORTED EVENT. THE RADIO FREQUENCY IDENTIFICATION (RFID) FRONT PANEL PRINTED CIRCUIT BOARD (PCB) WAS REPLACED TO ADDRESS THE ISSUE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A NON-CONFORMANCE BASED REVIEW OF THE SERIAL NUMBER WAS PERFORMED AND DID NOT REVEAL ANY POTENTIAL CONTRIBUTING FACTORS TO THE REPORTED COMPLAINT. A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. A REVIEW FOR COMPLAINTS REPORTED AGAINST THIS SERIAL NUMBER WAS PERFORMED. NO SIMILAR COMPLAINTS WERE REPORTED FOR THE PRODUCT SERIAL UNDER INVESTIGATION. THE ROOT CAUSE OF THE REPORTED EVENT IS ATTRIBUTED TO THE NONCONFORMING RFID PCB. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A CUSTOMER REPORTED THAT AN OPHTHALMIC CONSOLE EXHIBITED LOW LASER POWER. THE DETAILS ABOUT PROCEDURE AND PATIENT CONTACT WAS NOT REPORTED. ADDITIONAL INFORMATION HAD BEEN REQUESTED BUT NONE RECEIVED TILL DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 3297 | PUREPOINT LASER | LASER, OPHTHALMIC | HQF | ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER | NA | NA | 00380657505975 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |