FDA Adverse Event Malfunction Summary report: N

DEPTH STOP ADAPTER

MDR report key: 1659134 · Received April 13, 2010

Report

Report Number
3005677147-2010-00002
Event Type
Malfunction
Date Received
April 13, 2010
Date of Event
March 9, 2010
Report Date
March 9, 2010
Manufacturer
FHC, INC.
Product Code
HAW
PMA / PMN Number
K092562
Removal / Correction Number
3005677147-4/09/10-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION OF THE PRODUCT WAS PERFORMED BY FHC TECH SUPPORT PERSONNEL AT THE SURGERY. THIS EVALUATION FOUND THAT THE SCREW USED TO FIXATE THE LEAD WITHIN THE DBS DEPTH STOP ADAPTER HAD DAMAGED THE LEAD. THIS WAS A VALIDATION CASE FOR A RECENT PRODUCT IMPROVEMENT TO THE DEPTH STOP ADAPTER. A SPECIFIC WARNING IN THE DFU EXISTS FOR THIS DEVICE STATING THAT OVERTIGHTENING OF THE SCREW MAY DAMAGE THE LEAD. IMPROVEMENT WAS ALSO INSTALLED ON OTHER DEPTH STOP ADAPTERS. SUBSEQUENT FIELD ACTION IS UNDERWAY FOR THIS PRODUCT.

Description of Event or Problem · 1

TECH SUPPORT PERSONNEL ATTENDED A SURGERY WITH THE STAR DRIVE SYSTEM IN USE FOR PLACEMENT OF A DBS LEAD. SURGEON WAS VALIDATING AN IMPROVEMENT TO THE DEPTH STOP ADAPTER. THE DEPTH STOP ADAPTER USED TO HOLD THE DBS LEAD PRIOR TO PLACEMENT DAMAGED THE LEAD - THERE WAS A SHORT BETWEEN TWO OF THE LEAD CONTACTS. THIS WAS NOTED PRIOR TO LEAD PLACEMENT. A NEW LEAD WAS USED AND PLACED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEPTH STOP ADAPTER HAW FHC, INC.

Patients

Seq Age Sex Outcome Treatment
1 STAR DRIVE SYSTEM