FDA Adverse Event Malfunction Summary report: N

BD ULTRA-FINE¿ 4MM PEN NEEDLES

MDR report key: 16588681 · Received March 22, 2023

Report

Report Number
9616656-2023-00261
Event Type
Malfunction
Date Received
March 22, 2023
Date of Event
March 2, 2023
Report Date
May 11, 2023
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
UDI-DI
00382903204779
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 24-APR-2023. INVESTIGATION SUMMARY: ONE OPEN 32G X 4MM PEN NEEDLE SAMPLE WAS RETURNED FROM LOT. NO. 2242750, CAT. NO. 320477. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLE AND A BENT NON PATIENT END OF CANNULA WAS OBSERVED. DUE TO THE CONDITION THE SAMPLE WAS RETURNED NO CLOG TEST COULD BE CARRIED OUT. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS THE SAMPLE RETURNED WAS OPEN IT IS NOT POSSIBLE TO DETERMINE ROOT CAUSE.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD ULTRA-FINE¿ 4MM PEN NEEDLES THE MEDICATION COULD NOT BE DELIVERED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I HAVE ANOTHER BLOCKING NEEDLE WHICH I WILL KEEP SEPARATE.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD ULTRA-FINE¿ 4MM PEN NEEDLES THE MEDICATION COULD NOT BE DELIVERED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I HAVE ANOTHER BLOCKING NEEDLE WHICH I WILL KEEP SEPARATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
825937 BD ULTRA-FINE¿ 4MM PEN NEEDLES HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 2242750 00382903204779

Patients

Seq Age Sex Outcome Treatment
1 Unknown