FDA Adverse Event Malfunction Summary report: N

MRIDIIUM SERIES 1000 INFUSION SET

MDR report key: 16585672 · Received March 21, 2023

Report

Report Number
3005053560-2023-00006
Event Type
Malfunction
Date Received
March 21, 2023
Report Date
March 21, 2023
Manufacturer
IRADIMED CORPORATION
Product Code
FPK
PMA / PMN Number
K143369
Removal / Correction Number
EVENT 91846
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO PATIENT INJURY HAS BEEN REPORTED AS PART OF THE USE OF THE DEFECTIVE DEVICE. THIS IS ONE OF SEVERAL COMPLAINTS THAT RESULTED IN A CAPA INVESTIGATION INTO THE ISSUE. AFTER REVIEWING 1057 IV SETS RETURNED FROM CUSTOMERS WITH THE COMPLAINTS, IT WAS DETERMINED THAT THE SYRINGE CONNECTOR VENT WAS BLOCKED. IN ALL THE REPORTED COMPLAINTS, THE CUSTOMER INDICATED THERE WAS NO PATIENT INJURY, BUT THAT AN INLET OCCLUSION ALARM HAD OCCURRED RESULTING IN THE NEED TO REPLACE THE INFUSION SET. A HEALTH HAZARD EVALUATION (HHE) WAS PERFORMED AS PART OF THE INVESTIGATION INTO THE REPORTED EVENTS, AND ALTHOUGH THERE HAVE BEEN NO REPORTS OF INJURY, UNDER CERTAIN SEQUENCE OF EVENTS, THE HAZARDOUS SITUATIONS (UNDERDOSE AND/OR DELAY OF THERAPY) COULD LEAD TO POSSIBLE HARMS, INCLUDING SERIOUS ADVERSE HEALTH CONSEQUENCES. THE HHE DETERMINED THAT THE SEQUENCE OF EVENTS THAT COULD LEAD TO THE HAZARDOUS SITUATION HAS A REMOTE PROBABILITY OF OCCURRING, BUT DUE TO THE LEVEL OF SEVERITY IF IT WERE TO OCCUR, IT WAS DETERMINED THAT UPDATES TO THE PRODUCT LABELING AS PART OF A PRODUCT FIELD CORRECTION INFORMING THE USER TO VERIFY PROPER VENTING OF THE 1057 SET DURING THE INFUSION SET PRIMING PROCESS WAS THE APPROPRIATE ACTION. PRIOR TO THE HHE, THE LEVEL OF RISK OF AN INLET OCCLUSION WAS NOT EXPECTED TO RESULT IN SERIOUS ADVERSE HEALTH CONSEQUENCES AS THE SYSTEMS MITIGATIONS INCLUDE PROVIDING VISUAL AND AUDIBLE ALARMS FOR INLET OCCLUSIONS. HOWEVER, IN CONSIDERATION OF THE HHE'S FINDING THAT SERIOUS ADVERSE HEALTH CONSEQUENCES COULD OCCUR IN REMOTE CASES, THE RISK MANAGEMENT FILE WAS REVISED WITH A HIGHER SEVERITY LEVEL, THEREBY REQUIRING MEDICAL DEVICE REPORTING PER 21 CFR PART 803 DUE TO THE POTENTIAL OF HARM THAT MAY OCCUR. THE HHE WAS COMPLETED ON (B)(6) 2023 AND THIS IS CONSIDERED THE DATE THAT THE RISK WAS RE-EVALUATED RESULTING IN A "BECOME AWARE DATE" THEREBY TRIGGERING THIS MDR REPORT OF A PRODUCT DEFECT.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT THE SYRINGE VENT IS PLUGGED ON A SET FROM A 1057 SET, LOT 22L02N.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11748 MRIDIIUM SERIES 1000 INFUSION SET 1057 MRI SYRING ADPATER SET FPK IRADIMED CORPORATION 1057 22L02N

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other