BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD
Report
- Report Number
- 1119779-2023-00311
- Event Type
- Malfunction
- Date Received
- March 21, 2023
- Date of Event
- March 7, 2023
- Report Date
- March 31, 2023
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JSJ
- UDI-DI
- 00382904436247
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY: THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR RECENT COMPLAINT (B)(4) REGARDING BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254072 AND LOT NUMBER 3024190. EVENT DESCRIPTION: IT WAS REPORTED THAT ON CNA AGAR LOT NUMBER 3024190 THERE IS NO CULTURE AT ALL. (THE AGAR IS WELL SEEDED, CHECK MADE, AND THE BACTERIA FOUND ON THE COS ARE SUPPOSED TO GROW ON THIS AGAR TOO). COMPLAINT HISTORY REVIEW: THE COMPLAINTS TRENDS WERE REVIEWED. NO SIMILAR COMPLAINTS WERE RECEIVED FOR THE SAME CATALOG NUMBER. A TREND WAS NOT IDENTIFIED, THIS COMPLAINT WILL BE TREATED AS AN ISOLATED INCIDENT. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY RECORD WAS REVIEWED. NO DEVIATIONS WERE REGISTERED FROM VALIDATED MANUFACTURING PROCESSES. ALL RELEASE TESTING WAS SATISFACTORY. SAMPLE ANALYSIS: RETAIN SAMPLES WERE ANALYZED FOR THE GROWTH PROMOTING ABILITY OF THE MEDIUM. STRAINS USED FOR QUALITY RELEASE TESTING OF CATALOG NUMBER 254072 AND RECOMMENDED FOR USER QUALITY CONTROL OF THE MEDIUM WERE APPLIED TO MEDIUM OF LOT NO. 3024190. THE GROWTH OF SAID ORGANISMS WAS EXCELLENT. RETURN SAMPLES WERE NOT SHARED HOWEVER A PICTURE ILLUSTRATING ONE PLATES WITHOUT BACTERIAL GROWTH WAS SHARED AND A SECOND COS PLATE WITH BACTERIAL GROWTH (LIKE STAPHYLOCOCCUS SP.). EVALUATION RESULTS: AT THIS STAGE OF OUR INVESTIGATION, SYSTEMIC FAILURES OF THE VALIDATED MANUFACTURING PROCESSES CAN BE EXCLUDED. QC RELEASE TESTING OF THE AFFECTED LOT WAS SATISFACTORY. ANALYSIS OF THE COMPLAINT HISTORY SHOWED NO TRENDING OR VIOLATION OF ACTION LIMITS. INVESTIGATION CONCLUSION: BASED ON THE RESULTS OF THE INVESTIGATION, THE COMPLAINT CANNOT BE CONFIRMED. AS NO FURTHER COMPLAINTS REGARDING THIS LOT WERE NOTICED AND ALL ANALYSIS DONE BY THE QC DEPARTMENT YIELDED SATISFACTORY RESULTS, FURTHER INVESTIGATION WAS OMITTED. A DEFINITE ROOT CAUSE COULD NOT BE FOUND. HOWEVER, BD WILL CONTINUE TO MONITOR INCOMING COMPLAINTS FOR SIMILAR FAILURE TYPES. WE ARE SORRY FOR THE INCONVENIENCE.
DATE OF EVENT IS UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS. COMMON DEVICE NAME: CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US, BUT IT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 221353 WITH 510K # PREAMENDMENT. UNIQUE IDENTIFIER (UDI) #: (B)(4). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING THE BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT THERE WAS A PERFORMANCE ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE AGAIN PROBLEMS WITH YOUR CNA AGARS REF (B)(4). IN ATTACHMENT 2 PHOTOS OF THE SAME FILE: - COS AGAR WHERE THERE ARE MANY BACTERIA. - CNA AGAR WHERE THERE IS NO CULTURE AT ALL. (THE AGAR IS WELL SEEDED, CHECK MADE, AND THE BACTERIA FOUND ON THE COS ARE SUPPOSED TO GROW ON THIS AGAR TOO). OTHER AGARS ARE AFFECTED. LOT NUMBER 3024190.
IT WAS REPORTED THAT WHILE USING THE BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD THAT THERE WAS A PERFORMANCE ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE AGAIN PROBLEMS WITH YOUR CNA AGARS REF 254072. IN ATTACHMENT 2 PHOTOS OF THE SAME FILE: COS AGAR WHERE THERE ARE MANY BACTERIA. CNA AGAR WHERE THERE IS NO CULTURE AT ALL. (THE AGAR IS WELL SEEDED, CHECK MADE, AND THE BACTERIA FOUND ON THE COS ARE SUPPOSED TO GROW ON THIS AGAR TOO). OTHER AGARS ARE AFFECTED. LOT NUMBER 3024190.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11555 | BD COLUMBIA CNA AGAR WITH 5% SHEEP BLOOD | SEE H.10 | JSJ | BECTON, DICKINSON & CO. (SPARKS) | 443624 | 3024190 | 00382904436247 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |