FDA Adverse Event
Injury
Summary report: N
PITT EASY ABUTMENT
MDR report key: 1658194
·
Received April 13, 2010
Report
- Report Number
- 9611820-2010-00005
- Event Type
- Injury
- Date Received
- April 13, 2010
- Date of Event
- June 28, 2007
- Report Date
- March 15, 2010
- Manufacturer
- SYBRON IMPLANT SOLUTIONS GMBH
- Product Code
- NHA
- PMA / PMN Number
- K053242
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THAT A PATIENT HAD A FRACTURED ABUTMENT. THE PRODUCT WAS RETURNED TO SYBRON IMPLANT SOLUTIONS (B) (4) FOR EVALUATION. THE EVALUATION INDICATED THAT THE ABUTMENT MEETS PRODUCT SPECIFICATIONS AND THAT THE FRACTURE WAS CAUSED BY LOADING AFTER LOOSENING. THIS IS NOT A PRODUCT FAILURE.
Description of Event or Problem · 1
ON MARCH 15, 2010 A DOCTOR REPORTED TO SYBRON IMPLANT SOLUTIONS (B) (4) THAT A PATIENT HAD A PITT EASY IMPLANT ABUTMENT SCREW FRACTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PITT EASY ABUTMENT | ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS | NHA | SYBRON IMPLANT SOLUTIONS GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other| R |