VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE
Report
- Report Number
- 9610773-2023-00811
- Event Type
- Malfunction
- Date Received
- March 20, 2023
- Date of Event
- January 12, 2023
- Report Date
- April 20, 2023
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HET
- UDI-DI
- 04042761074971
- PMA / PMN Number
- K111788
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS MEDWATCH IS BEING SUPPLEMENTED WITH ADDITIONAL INFORMATION OBTAINED AND THE MANUFACTURER'S FINAL INVESTIGATION RESULTS. DURING THEIR INSPECTION, THE SBC DETECTED VARYING-COLORED IMAGES (PURPLE/GREEN). BY DISASSEMBLING THE DEVICE AND TESTING THE COMPONENTS INDIVIDUALLY, THE IMAGE ERROR COULD BE TRACED BACK TO THE R-UNIT. FURTHERMORE, THE SBC FOUND THE OUTER TUBE TO BE DENTED. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER WITHOUT SHOWING ANY NON-CONFORMITIES OR DEVIATIONS REGARDING THE DESCRIBED ISSUES.
THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND REPAIR. THE INVESTIGATION IS STILL IN PROGRESS; THEREFORE, THE ROOT CAUSE OF THE REPORTED DEFECT CANNOT BE DETERMINED AT THIS TIME. HOWEVER, IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED ACCORDINGLY. IN GENERAL, THE CUSTOMER IS REQUIRED TO INSPECT THE DEVICE FOR DEFECTS, CHECK THE FUNCTION OF ALL DEVICES AND HAVE ALTERNATE EQUIPMENT PRIOR TO USE. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE. ADDITIONAL PMA/510(K): K190744.
IT WAS REPORTED TO OLYMPUS THAT ON (B)(6) 2023, A VIDEO TELESCOPE, DURING PREPARATION FOR USE, THE IMAGE WAS GREEN. THEN CHANGED TO ANOTHER DEVICE AND FINISHED THE PROCEDURE. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT REPORTED TO OLYMPUS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1650312 | VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE | LAPAROSCOPE | HET | OLYMPUS WINTER & IBE GMBH | WA50042A | 04042761074971 |