CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2023-01547
- Event Type
- Malfunction
- Date Received
- March 20, 2023
- Date of Event
- March 9, 2023
- Report Date
- September 20, 2024
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- 003
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
THE GETINGE SERVICE TERRITORY MANAGER (STM) THAT DISCOVERED THE SPILL REPLACED THE POWER MANAGEMENT BOARD. COMPLETED PM WITH FULL CALIBRATION, FUNCTIONAL TESTING AND SAFETY CHECKS TO FACTORY SPECIFICATIONS. UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS PER FACTORY SPECIFICATIONS. THE UNIT WAS APPROVED FOR CLINICAL USE AND RETURNED TO THE DEPARTMENT. THE FAILURE ANALYSIS AND TESTING DEPARTMENT RECEIVED A POWER MANAGEMENT BOARD P/N 0670-00-1162, S/N (B)(6), WITH A REPORTED OF THE ¿SALINE SPILL ON UNIT, DAMAGED BOARD¿. PERFORMED VISUAL INSPECTION OF THE POWER MANAGEMENT BOARD PER THE CARDIOSAVE SERVICE MANUAL AND FOUND NO VISUAL DAMAGE AND THE PART LOOKS TO BE IN GOOD CONDITION. INSTALLED POWER MANAGEMENT BOARD INTO IABP CARDIOSAVE TEST FIXTURE FOR TESTING OF THE REPORTED PROBLEM. PERFORMED AND TESTED THE SYSTEM (45 MINUTES) IN ACCORDANCE WITH PROCEDURE AND THE CARDIOSAVE SERVICE MANUAL IN AN ATTEMPT TO RECREATE THE REPORTED PROBLEM. FOUND THAT ALL FUNCTIONS WERE NORMAL. THE TESTS DID NOT TRIGGER THE REPORTED PROBLEM OF ¿SALINE SPILL ON UNIT, DAMAGED BOARD¿. THE FAILURE ANALYSIS AND TESTING DEPARTMENT WAS UNABLE TO REPLICATE THE FAILURE EXPERIENCED BY THE CUSTOMER. SENDING THE POWER MANAGEMENT BOARD TO THE SUPPLIER FOR FURTHER FAILURE ANALYSIS. THE PART ASSOCIATED TO THIS COMPLAINT WAS SUPPLIED TO US BY THE SUPPLIER (B)(4), AND IT IS A REVISION A POWER MANAGEMENT BOARD. OUR NEW SUPPLIER FOR THE POWER MANAGEMENT BOARD IS (B)(4), AND THEY ARE ONLY ABLE TO TEST 0670-00-1162 BOARDS THAT ARE REVISION G AND NEWER. THE PART NUMBER 0670-00-1162, POWER MANAGEMENT, ROHS SERIAL NUMBER (B)(6) RECEIVED FOR THIS COMPLAINT WILL NOT BE GOING BACK TO THE SUPPLIER. THE FAT DEPT WILL RETAIN THIS PART AS PER PROCEDURE.
IT WAS REPORTED THAT DURING PREVENTIVE MAINTENANCE (PM) PERFORMED BY A GETINGE SERVICE TERRITORY MANAGER (STM) , THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) IT WAS DISCOVERED THAT SALINE HAD BEEN SPILLED ON THE UNIT. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 988122 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |