FDA Adverse Event Malfunction Summary report: N

IMPLANTABLE COLLAMER LENS (ICL)

MDR report key: 16567677 · Received March 17, 2023

Report

Report Number
2023826-2023-00940
Event Type
Malfunction
Date Received
March 17, 2023
Date of Event
February 7, 2023
Report Date
February 20, 2023
Manufacturer
STAAR SURGICAL COMPANY
Product Code
QCB
UDI-DI
00841542121187
PMA / PMN Number
P030016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTED DATA: B5- THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.2 VTICMO 13.2 IMPLANTABLE COLLAMER LENS OF DIOPTER -16.50/+3.50/093 (SPHERE/CYLINDER/AXIS) INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. THE PATIENT EXPERIENCED AN EXCESSIVE VAULT, BLURRED VISION, AND RESIDUAL REFRACTIVE ERROR. REPORTEDLY " PATIENT HAS A HIGH VAULT OU BUT THERE HAVE BEEN NO ISSUES WITH IOP. HIS PATIENT ALSO HAS RESIDUAL REFRACTIVE ERROR OD GREATER THAN OS. HOWEVER, THE PATIENT WAS AT UPPER EXTREME OF MYOPIA OD VS OS. HE IS NOT COMPLAINING BUT WANTED TO RECEIVE SOME ADVICE ON FOLLOWING AND OPTIONS FOR THE PATIENT." THE LENS REMAINS IMPLANTED. THE PROBLEM WAS NOT RESOLVED. STATUS OF THE EYE: " BLURRY VISION". THE CAUSE OF THE EVENT IS UNKNOWN. CAUSE DETAILS- "HIGH VAULT". CLAIM# (B)(4).

Additional Manufacturer Narrative · 0

WORK ORDER SEARCH: NO SIMILAR COMPLAINT WAS REPORTED FOR UNITS WITHIN THE SAME LOT. CLAIM # (B)(4).

Description of Event or Problem · 0

THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.2MM VTICMO13.2 IMPLANTABLE COLLAMER LENS, -16.50/+3.5/093 (SPHERE/CYLINDER/AXIS), INTO THE PATIENTS LEFT EYE (OS) ON (B)(6) 2023. THE LENS WAS REPORTED AS EXCESSIVE VAULTING AND THE PATIENT COMPLAINED OF BLURRY VISION. THE LENS REMAINED IMPLANTED. THE CAUSE OF THE EVENT WAS UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666044 IMPLANTABLE COLLAMER LENS (ICL) PHAKIC TORIC INTRAOCULAR LENS QCB STAAR SURGICAL COMPANY VTICMO13.2 NA 00841542121187

Patients

Seq Age Sex Outcome Treatment
1 38 YR Female UNK.