BD MICRO-FINE¿ PRO PEN NEEDLE
Report
- Report Number
- 9616656-2023-00233
- Event Type
- Malfunction
- Date Received
- March 17, 2023
- Date of Event
- February 21, 2023
- Report Date
- June 15, 2023
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY THREE OPEN 32G X 4MM PEN NEEDLE SAMPLE WERE RETURNED FROM LOT. NO. 2040131, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON ALL THREE SAMPLES AND BENT NON PATIENT END OF CANNULA WAS OBSERVED ON TWO SAMPLES AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED ON THE REMAINING SAMPLE. DUE TO THE CONDITION THE SAMPLES WERE RETURNED NO CLOG TEST COULD BE CARRIED OUT. TWO OPEN 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 2060476, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLES AND A BENT NON PATIENT END OF CANNULA WAS OBSERVED ON ONE SAMPLE. A CLOG TEST AS PER Q-SOP-183-DL WAS CARRIED OUT ON THE REMAINING SAMPLE AND NO ISSUES WERE OBSERVED. TWO OPEN 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 1195507, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLES AND A BENT NON PATIENT END OF CANNULA WAS OBSERVED ON ONE SAMPLE AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED ON THE SECOND SAMPLE. DUE TO THE CONDITION THE SAMPLES WERE RETURNED NO CLOG TEST COULD BE CARRIED OUT. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS ALL SAMPLES RETURNED WERE OPEN IT IS NOT POSSIBLE TO DETERMINE ROOT CAUSE.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 2040131; MEDICAL DEVICE EXPIRATION DATE: 28-FEB-2027; DEVICE MANUFACTURE DATE: 09-FEB-2022. MEDICAL DEVICE LOT #: 1195507; MEDICAL DEVICE EXPIRATION DATE: 31-JUL-2026; DEVICE MANUFACTURE DATE: 14-JUL-2021. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT 1 BD MICRO-FINE¿ PRO PEN NEEDLE EACH FROM LOTS 1195507 AND 2040131 WERE FOUND WITH BROKEN NON-PATIENT ENDS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE BDJ INVESTIGATION TEAM: "BDJ CONFIRMED, 2 NP NEEDLE BROKEN."
IT WAS REPORTED THAT 1 BD MICRO-FINE¿ PRO PEN NEEDLE EACH FROM LOTS 1195507 AND 2040131 WERE FOUND WITH BROKEN NON-PATIENT ENDS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE BDJ INVESTIGATION TEAM: "BDJ CONFIRMED... 2 NP NEEDLE BROKEN".