FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE¿ PRO PEN NEEDLE

MDR report key: 16566407 · Received March 17, 2023

Report

Report Number
9616656-2023-00233
Event Type
Malfunction
Date Received
March 17, 2023
Date of Event
February 21, 2023
Report Date
June 15, 2023
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY THREE OPEN 32G X 4MM PEN NEEDLE SAMPLE WERE RETURNED FROM LOT. NO. 2040131, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON ALL THREE SAMPLES AND BENT NON PATIENT END OF CANNULA WAS OBSERVED ON TWO SAMPLES AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED ON THE REMAINING SAMPLE. DUE TO THE CONDITION THE SAMPLES WERE RETURNED NO CLOG TEST COULD BE CARRIED OUT. TWO OPEN 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 2060476, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLES AND A BENT NON PATIENT END OF CANNULA WAS OBSERVED ON ONE SAMPLE. A CLOG TEST AS PER Q-SOP-183-DL WAS CARRIED OUT ON THE REMAINING SAMPLE AND NO ISSUES WERE OBSERVED. TWO OPEN 32G X 4MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 1195507, CAT. NO. 320559. VISUAL EXAMINATION WAS CARRIED OUT ON THE RETURNED SAMPLES AND A BENT NON PATIENT END OF CANNULA WAS OBSERVED ON ONE SAMPLE AND A BROKEN NON PATIENT END OF CANNULA WAS OBSERVED ON THE SECOND SAMPLE. DUE TO THE CONDITION THE SAMPLES WERE RETURNED NO CLOG TEST COULD BE CARRIED OUT. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS ALL SAMPLES RETURNED WERE OPEN IT IS NOT POSSIBLE TO DETERMINE ROOT CAUSE.

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 2040131; MEDICAL DEVICE EXPIRATION DATE: 28-FEB-2027; DEVICE MANUFACTURE DATE: 09-FEB-2022. MEDICAL DEVICE LOT #: 1195507; MEDICAL DEVICE EXPIRATION DATE: 31-JUL-2026; DEVICE MANUFACTURE DATE: 14-JUL-2021. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD MICRO-FINE¿ PRO PEN NEEDLE EACH FROM LOTS 1195507 AND 2040131 WERE FOUND WITH BROKEN NON-PATIENT ENDS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE BDJ INVESTIGATION TEAM: "BDJ CONFIRMED, 2 NP NEEDLE BROKEN."

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD MICRO-FINE¿ PRO PEN NEEDLE EACH FROM LOTS 1195507 AND 2040131 WERE FOUND WITH BROKEN NON-PATIENT ENDS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE BDJ INVESTIGATION TEAM: "BDJ CONFIRMED... 2 NP NEEDLE BROKEN".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
665935 BD MICRO-FINE¿ PRO PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. SEE H10

Patients

Seq Age Sex Outcome Treatment
1 Unknown