MDR report key: 16564 · Received June 15, 1994

Report

Report Number
MW4000490
Date Received
June 15, 1994
Report Date
June 15, 1994
Manufacturer
MARQUETTE ELECTRONICS, INC.
Product Code
DPS
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

EKG MACHINE WAS FOUND TO LOCK UP WHILE MONITORING A PT THAT HAD A DUAL CHAMBER PACING MODE. PACEMAKER IMPLANTED. THE PACEMAKER WAS OPERATED IN THE PT DEPENDENT MODE AT A RATE OF 100 BPM, PSA.6V, SA VENT .7MA, VENT 774 OHMS, VENT SLOW RATE 3.17V, VENT PEAK 22.2 MM V, FSA ATR .8 MA, ATR RES 866 OHMS, P-WAVE AMP 2.0 MV, F-WAVE AMP 1.8 MV. THE EKG MACHINE SEEMS TO HAVE SOME SORT OF SOFTWARE GLITCH WHICH CAUSES THE MACHINE TO LOCK UP UNDER THESE CONDITIONS. THE MFR WAS NOTIFIED OF THE PROBLEM. NO PT INCIDENT WAS REPORTED SINCE THIS DID NOT CAUSE ANY ADVERSE PT PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAC VU EKG MACHINE DPS MARQUETTE ELECTRONICS, INC. MAC VU

Patients

Seq Age Sex Outcome Treatment
1 *