CERTAIN¿ UCLA CASTABLE NON-HEXED ABUTMENT 3.4MM(D) W/LARGE DIA. TI SCREW
Report
- Report Number
- 0001038806-2023-00520
- Event Type
- Malfunction
- Date Received
- March 16, 2023
- Date of Event
- December 11, 2022
- Report Date
- November 2, 2023
- Manufacturer
- BIOMET 3I
- Product Code
- NHA
- UDI-DI
- 00844868006558
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED TWO (2) IMUCC2T, (CERTAIN UCLA CASTABLE NON-HEXED ABUTMENT 3.4MM(D) W/LARGE DIA. TI SCREW) FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED SIGNS OF USE, BOTH SCREWS WERE FRACTURED AT THE THREADS. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1212548. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1212548 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: FRACTURE SCREW. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION IS ATTRIBUTED TO PATIENT FACTORS DUE TO PARAFUNCTIONAL HABITS (BRUXISM) OVER THE IMPLANTATION PERIOD. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR, AND THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT WEIGHT UNKNOWN / NOT PROVIDED. ADDITIONAL DEVICE INFORMATION UNKNOWN / NOT PROVIDED. PREMARKET IDENTIFICATION UNKNOWN / NOT PROVIDED. DEVICE MANUFACTURER DATE UNKNOWN / NOT PROVIDED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
THE DOCTOR REPORTS THAT THE TWO SCREWS HAVE FRACTURED THROUGH/BY THE BODY. THE AFFECTED DENTAL POSITIONS ARE #43 AND #46. THE DOCTOR INDICATES IN THE PER THAT THE PROCEDURE WAS CONCLUDED WITH THE PLACEMENT OF ANOTHER IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 455863 | CERTAIN¿ UCLA CASTABLE NON-HEXED ABUTMENT 3.4MM(D) W/LARGE DIA. TI SCREW | DENTAL ABUTMENT | NHA | BIOMET 3I | 1212548 | 00844868006558 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Prefer Not To Disclose |