FDA Adverse Event Malfunction Summary report: N

WAVE SIDE GRASPER FORCEPS

MDR report key: 1655135 · Received April 7, 2010

Report

Report Number
1418479-2010-00005
Event Type
Malfunction
Date Received
April 7, 2010
Date of Event
February 22, 2010
Report Date
April 7, 2010
Manufacturer
RICHARD WOLF MEDICAL INSTRUMENTS
Product Code
BWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

JAW BROKEN. EVALUATION SUMMARY: VISUAL INSPECTION OF THE FORCEPS WITH BROKEN JAW SHOWED NO CORROSION AT THE POINT OF BREAKAGE. THE FORCEPS IS APPROXIMATELY 6 YEARS OLD. THE INSULATION IS IN GOOD CONDITION. OUR REPAIR RECORDS SHOW THAT THIS FACILITY PURCHASED A MORE THAN ONE OF THESE FORCEPS OF THIS LOT NUMBER 1U03 IN 2004. WE CONSIDER THIS AN ISOLATED OCCURRENCE SINCE NO OTHER COMPLAINTS OF THIS TYPE ARE IN THE SYSTEM. CAUSE OF EVENT: NORMAL WEAR AND TEAR FOR THIS PERIOD OF TIME IN USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE GRASPER DURING A LAPAROSCOPIC SIGMOID RESECTION, THE JAW BROKE OFF INSIDE THE PATIENT. THE PIECE WAS RETRIEVED AND ACCOUNTED FOR BY THE SURGEON AND THE OR STAFF. THE JAW PIECE WAS DISCARDED. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WAVE SIDE GRASPER FORCEPS WAVE SIDE GRASPER FORCEPS BWB RICHARD WOLF MEDICAL INSTRUMENTS 8383.2937 1U03

Patients

Seq Age Sex Outcome Treatment
1 48 YR