FDA Adverse Event Malfunction Summary report: N

BD INTIMA II¿ IV CATHETER PRN ADAPTER

MDR report key: 16550850 · Received March 15, 2023

Report

Report Number
3014704491-2023-00056
Event Type
Malfunction
Date Received
March 15, 2023
Date of Event
February 11, 2023
Report Date
March 27, 2023
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9050858. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA II¿ IV CATHETER PRN ADAPTER TUBING RUPTURED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "DURING ENHANCED CT TO CREATE A CONTRAST AGENT, THE Y-SHAPED TUBE OF THE VENOUS INDWELLING NEEDLE PARTIALLY RUPTURED."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA II¿ IV CATHETER PRN ADAPTER TUBING RUPTURED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "DURING ENHANCED CT TO CREATE A CONTRAST AGENT, THE Y-SHAPED TUBE OF THE VENOUS INDWELLING NEEDLE PARTIALLY RUPTURED.".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
450925 BD INTIMA II¿ IV CATHETER PRN ADAPTER INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 0231196

Patients

Seq Age Sex Outcome Treatment
1 Unknown