FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 16548198 · Received March 15, 2023

Report

Report Number
1221359-2023-00504
Event Type
Malfunction
Date Received
March 15, 2023
Date of Event
January 1, 2023
Report Date
May 23, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DATE PROVIDED IN THE EVENT IS AN ESTIMATE AS THE EXACT DATE OF OCCURRENCE WAS NOT PROVIDED. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. SINGLE USE, DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

THE DATE PROVIDED IN B3 IS AN ESTIMATE AS THE EXACT DATE OF OCCURRENCE WAS NOT PROVIDED. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 212527 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 212527, TEST BASE PART NUMBER 195-430WL/ LOT: 209979. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 212527 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Description of Event or Problem · 0

THE CONSUMER REPORTED SIX (6) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED BY HERSELF ON MULTIPLE DAYS. THIS MFR. REPORT ADDRESSES TEST SIX (6) OF SIX (6). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON AN UNKNOWN DATE ON A NASAL KITTED SWAB. ADDITIONAL TESTING WAS PERFORMED ON AN UNKNOWN DATE VIA FLOW FLEX RAPID ANTIGEN TEST GENERATING A POSITIVE RESULT. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT..

Description of Event or Problem · 0

THE CONSUMER REPORTED SIX (6) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED BY HERSELF ON MULTIPLE DAYS. THIS MFR. REPORT ADDRESSES TEST SIX (6) OF SIX (6). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON AN UNKNOWN DATE ON A NASAL KITTED SWAB. ADDITIONAL TESTING WAS PERFORMED ON AN UNKNOWN DATE VIA FLOW FLEX RAPID ANTIGEN TEST GENERATING A POSITIVE RESULT. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
758288 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 212527 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 39 YR Male