FDA Adverse Event Injury Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 1654620 · Received April 5, 2010

Report

Report Number
1722139-2010-00042
Event Type
Injury
Date Received
April 5, 2010
Date of Event
March 30, 2010
Report Date
March 31, 2010
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ACTUAL DEVICE WAS EVALUATED. RESULTS: THE EVALUATION INCLUDED PERFORMANCE TESTING (ACCURACY, VERIFICATION OF AIR/NO FOOD ALARM, ETC.). MULTIPLE FORMULAS AND SETTINGS WERE TO RECREATE THE FAILURE EXPERIENCED BY THE CUSTOMER. THE PUMP PERFORMED ACCORDING TO SPECIFICATION DURING EACH RUN (ALARMED AND SHUT OFF AFTER FOOD WAS GONE). CONCLUSION: THE ALLEGED COMPLAINT/DEFECT COULD NOT BE DUPLICATED. WE ARE ATTEMPTING TO CONTACT THE CUSTOMER/PT TO OBTAIN MORE INFORMATION. IF MORE CUSTOMER/PT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT MAY BE SUBMITTED.

Description of Event or Problem · 1

REPORTED BY THE CUSTOMER AS: PUMP CONTINUED TO PUMP AFTER FOOD WAS GONE. CUSTOMER STATED "PUMP KEEPS RUNNING WHEN BAG IS EMPTY." PT INJURY? NO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MEDICAL DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1