HEARTMATE II SYSTEM CONTROLLER, EU
Report
- Report Number
- 2916596-2010-00073
- Event Type
- Death
- Date Received
- April 7, 2010
- Date of Event
- March 21, 2010
- Report Date
- March 22, 2010
- Manufacturer
- THORATEC CORP.
- Product Code
- DSQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PATIENT EXPIRED ON (B) (6) 2010. SYSTEM CONTROLLERS, (B) (4) ARE NOT THE ORIGINAL EQUIPMENT FOR THIS PATIENT. THE TWO SYSTEM CONTROLLERS BELONGED TO A DIFFERENT PATIENT BEFORE BEING PROVIDED TO THIS PATIENT FOR USE. SYSTEM CONTROLLER (B) (4) WAS IN USE BY THE PATIENT AT THE TIME OF THE EVENT, PRIOR TO BEING EXCHANGED. AN ATTEMPT TO DOWNLOAD THE HISTORY OF THIS SYSTEM CONTROLLER WAS UNSUCCESSFUL AS THE SYSTEM CONTROLLER WOULD NOT COMMUNICATE WITH THE POWER MODULE. SYSTEM CONTROLLER (B) (4) WAS USED TO REPLACE (B) (4). THE HISTORY OF (B) (4) REVEALED APPROXIMATELY 17 HOURS OF CONTINUOUS FLOW ALARMS. THE MANUFACTURER IS ATTEMPTING TO ACQUIRE THE SYSTEM CONTROLLER AND EXPLANTED PUMP FOR ANALYSIS. ALL OTHER EQUIPMENT FOR THIS PATIENT WAS CONFISCATED BY THE POLICE DEPARTMENT AND WILL HELD UNTIL FURTHER NOTICE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETED.
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE. APPROXIMATELY 21 MONTHS POST-IMPLANT, THE PATIENT WAS AT HOME AND NOT FEELING WELL ALL DAY AND COLLAPSED IN THE EVENING WHILE CONNECTED TO BATTERIES. EMERGENCY SERVICE WAS CALLED BY THE PATIENT'S WIFE AND EMERGENCY SERVICES CONTACTED THE IMPLANTING HOSPITAL FOR ADVICE. THE SYSTEM CONTROLLER WAS EXCHANGED BASED ON THE DISCUSSION AND THE PATIENT WAS TRANSPORTED TO A NEARBY HOSPITAL. THE PATIENT WAS REPORTED TO BE UNCONSCIOUS DURING THE ENTIRE TIME AND EXPIRED APPROXIMATELY 17 HOURS AFTER THE SYSTEM CONTROLLER EXCHANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTMATE II SYSTEM CONTROLLER, EU | LVAD SYSTEM CONTROLLER | DSQ | THORATEC CORP. | 103692 | 84133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Death |