FDA Adverse Event Death Summary report: N

HEARTMATE II SYSTEM CONTROLLER, EU

MDR report key: 1654350 · Received April 7, 2010

Report

Report Number
2916596-2010-00073
Event Type
Death
Date Received
April 7, 2010
Date of Event
March 21, 2010
Report Date
March 22, 2010
Manufacturer
THORATEC CORP.
Product Code
DSQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT EXPIRED ON (B) (6) 2010. SYSTEM CONTROLLERS, (B) (4) ARE NOT THE ORIGINAL EQUIPMENT FOR THIS PATIENT. THE TWO SYSTEM CONTROLLERS BELONGED TO A DIFFERENT PATIENT BEFORE BEING PROVIDED TO THIS PATIENT FOR USE. SYSTEM CONTROLLER (B) (4) WAS IN USE BY THE PATIENT AT THE TIME OF THE EVENT, PRIOR TO BEING EXCHANGED. AN ATTEMPT TO DOWNLOAD THE HISTORY OF THIS SYSTEM CONTROLLER WAS UNSUCCESSFUL AS THE SYSTEM CONTROLLER WOULD NOT COMMUNICATE WITH THE POWER MODULE. SYSTEM CONTROLLER (B) (4) WAS USED TO REPLACE (B) (4). THE HISTORY OF (B) (4) REVEALED APPROXIMATELY 17 HOURS OF CONTINUOUS FLOW ALARMS. THE MANUFACTURER IS ATTEMPTING TO ACQUIRE THE SYSTEM CONTROLLER AND EXPLANTED PUMP FOR ANALYSIS. ALL OTHER EQUIPMENT FOR THIS PATIENT WAS CONFISCATED BY THE POLICE DEPARTMENT AND WILL HELD UNTIL FURTHER NOTICE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE. APPROXIMATELY 21 MONTHS POST-IMPLANT, THE PATIENT WAS AT HOME AND NOT FEELING WELL ALL DAY AND COLLAPSED IN THE EVENING WHILE CONNECTED TO BATTERIES. EMERGENCY SERVICE WAS CALLED BY THE PATIENT'S WIFE AND EMERGENCY SERVICES CONTACTED THE IMPLANTING HOSPITAL FOR ADVICE. THE SYSTEM CONTROLLER WAS EXCHANGED BASED ON THE DISCUSSION AND THE PATIENT WAS TRANSPORTED TO A NEARBY HOSPITAL. THE PATIENT WAS REPORTED TO BE UNCONSCIOUS DURING THE ENTIRE TIME AND EXPIRED APPROXIMATELY 17 HOURS AFTER THE SYSTEM CONTROLLER EXCHANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II SYSTEM CONTROLLER, EU LVAD SYSTEM CONTROLLER DSQ THORATEC CORP. 103692 84133

Patients

Seq Age Sex Outcome Treatment
1 67 YR Death