VERITAS ADVANCED INFUSION PACK
Report
- Report Number
- 3012236936-2023-00612
- Event Type
- Injury
- Date Received
- March 14, 2023
- Date of Event
- February 15, 2023
- Report Date
- May 3, 2023
- Manufacturer
- AMO MANUFACTURING USA, LLC
- Product Code
- HQC
- UDI-DI
- 35050474700902
- PMA / PMN Number
- K203060
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION: ADDITIONAL INFORMATION WAS RECEIVED, AND IT WAS LEARNT THAT HOSPITAL IS STRICT ABOUT THE CLEANING PROCESS AND HAS PERFORMED IN-HOUSE CULTURE TEST FOR ALL THE MATERIALS USED (INCLUDING PHACO PACK, OVD {OPHTHALMIC VISCOELASTIC DEVICES} AND BSS {BALANCED SALT SOLUTION} WITH SAME LOT NUMBER USED FOR ALL THESE SURGERIES, NOTHING WAS FOUND FROM THE CULTURE TEST SO NOT SURE IF IT WAS DUE TO PHACO HANDPIECE. REPORTEDLY IF IT WAS A CLEANING PROCESS ISSUE, SAME ISSUE MIGHT HAVE HAPPENED WITH PATIENTS WHO HAD CATARACT SURGERY IN THE SAME OPERATING ROOM AS DEVICE FROM ANOTHER MANUFACTURER WAS USED BY OTHER PHYSICIANS AFTERWARDS. BSS FROM ANOTHER MANUFACTURER WAS USED AND NEW BOTTLES WERE USED FOR EACH PATIENT. NO FURTHER INFORMATION WAS PROVIDED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
AGE OR DATE OF BIRTH, WEIGHT AND ETHNICITY: UNKNOWN/ NOT PROVIDED. PATIENT INFORMATION CANNOT BE PROVIDED DUE TO PERSONAL DATA PRIVACY LEGISLATION/POLICY. TELEPHONE NUMBER: UNKNOWN, INFORMATION NOT PROVIDED. DEVICE EVALUATION: THE PRODUCT TESTING COULD NOT BE PERFORMED AS THE PRODUCT WAS NOT RETURNED. THE REPORTED COMPLAINT CANNOT BE CONFIRMED. MANUFACTURING RECORD EVALUATION: BASED ON THE MANUFACTURING RECORDS REVIEW, AND HISTORICAL COMPLAINT REVIEW, THERE IS NO INDICATION OF A PRODUCT MALFUNCTION OR PRODUCT QUALITY DEFICIENCY. NO NONCONFORMITY REPORT, DOCUMENTATION OR LABELING CHANGES, AND ESCALATIONS ARE REQUIRED. CONCLUSION: AS A RESULT OF THE INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT PATIENT WHO HAD CATARACT SURGERY WITH VERITAS AI TUBING PACK WAS DIAGNOSED WITH ENDOPHTHALMITIS. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS 6 OF 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 914840 | VERITAS ADVANCED INFUSION PACK | UNIT, PHACOFRAGMENTATION | HQC | AMO MANUFACTURING USA, LLC | VRT-AI | 60416304 | 35050474700902 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |