FDA Adverse Event Injury Summary report: N

VERITAS ADVANCED INFUSION PACK

MDR report key: 16542246 · Received March 14, 2023

Report

Report Number
3012236936-2023-00612
Event Type
Injury
Date Received
March 14, 2023
Date of Event
February 15, 2023
Report Date
May 3, 2023
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
HQC
UDI-DI
35050474700902
PMA / PMN Number
K203060
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: ADDITIONAL INFORMATION WAS RECEIVED, AND IT WAS LEARNT THAT HOSPITAL IS STRICT ABOUT THE CLEANING PROCESS AND HAS PERFORMED IN-HOUSE CULTURE TEST FOR ALL THE MATERIALS USED (INCLUDING PHACO PACK, OVD {OPHTHALMIC VISCOELASTIC DEVICES} AND BSS {BALANCED SALT SOLUTION} WITH SAME LOT NUMBER USED FOR ALL THESE SURGERIES, NOTHING WAS FOUND FROM THE CULTURE TEST SO NOT SURE IF IT WAS DUE TO PHACO HANDPIECE. REPORTEDLY IF IT WAS A CLEANING PROCESS ISSUE, SAME ISSUE MIGHT HAVE HAPPENED WITH PATIENTS WHO HAD CATARACT SURGERY IN THE SAME OPERATING ROOM AS DEVICE FROM ANOTHER MANUFACTURER WAS USED BY OTHER PHYSICIANS AFTERWARDS. BSS FROM ANOTHER MANUFACTURER WAS USED AND NEW BOTTLES WERE USED FOR EACH PATIENT. NO FURTHER INFORMATION WAS PROVIDED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

AGE OR DATE OF BIRTH, WEIGHT AND ETHNICITY: UNKNOWN/ NOT PROVIDED. PATIENT INFORMATION CANNOT BE PROVIDED DUE TO PERSONAL DATA PRIVACY LEGISLATION/POLICY. TELEPHONE NUMBER: UNKNOWN, INFORMATION NOT PROVIDED. DEVICE EVALUATION: THE PRODUCT TESTING COULD NOT BE PERFORMED AS THE PRODUCT WAS NOT RETURNED. THE REPORTED COMPLAINT CANNOT BE CONFIRMED. MANUFACTURING RECORD EVALUATION: BASED ON THE MANUFACTURING RECORDS REVIEW, AND HISTORICAL COMPLAINT REVIEW, THERE IS NO INDICATION OF A PRODUCT MALFUNCTION OR PRODUCT QUALITY DEFICIENCY. NO NONCONFORMITY REPORT, DOCUMENTATION OR LABELING CHANGES, AND ESCALATIONS ARE REQUIRED. CONCLUSION: AS A RESULT OF THE INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT WHO HAD CATARACT SURGERY WITH VERITAS AI TUBING PACK WAS DIAGNOSED WITH ENDOPHTHALMITIS. NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS 6 OF 6.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
914840 VERITAS ADVANCED INFUSION PACK UNIT, PHACOFRAGMENTATION HQC AMO MANUFACTURING USA, LLC VRT-AI 60416304 35050474700902

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other