FDA Adverse Event Injury Summary report: N

LINEAR ST

MDR report key: 16542148 · Received March 14, 2023

Report

Report Number
3006630150-2023-01270
Event Type
Injury
Date Received
March 14, 2023
Date of Event
August 2, 2022
Report Date
March 14, 2023
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729767725
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS, UPN: M365SC2218500, MODEL: SC-2218-50, SERIAL: (B)(4), BATCH: 7103483.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS NOT GETTING AN ADEQUATE PAIN RELIEF DESPITE REPROGRAMMING DONE DUE TO LEAD MIGRATION. THIS WAS CONFIRMED THROUGH AN X-RAY. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE LEAD WAS REPOSITIONED. THE PATIENT WAS DOING WELL POSTOPERATIVELY. NOTHING WAS ADDED OR REMOVED DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
914834 LINEAR ST STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2218-50 7103407 08714729767725

Patients

Seq Age Sex Outcome Treatment
1 49 YR Female Required Intervention