INRATIO
Report
- Report Number
- 2027969-2010-00381
- Event Type
- Malfunction
- Date Received
- March 31, 2010
- Date of Event
- March 10, 2010
- Report Date
- March 31, 2010
- Manufacturer
- BIOSITE INCORPORATED
- Product Code
- JPA
- PMA / PMN Number
- K021923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B) (6) 2010, INRATIO: 2.8, REFERENCE: 1.2, MEAN: 2.00, CONFIDENCE LIMITS: 1.3-2.7. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES FALL OUTSIDE THE LIMITS, SO THE CRITERIA ARE NOT MET AND THE VALUES ARE DISCREPANT BEYOND THE DOCUMENTED VARIABILITY FOR PT/INR TESTING. THIS WOULD BE SUBJECT TO STRIP ACCURACY TESTING AS PART OF THE COMPLAINT INVESTIGATION. A 2.6, 2.2, 1.6 AND 1.5 INR ARE EXCLUDED FROM COMPARISON TESTS SINCE IT WAS PERFORMED ON DIFFERENT DAYS. SINCE THE TIME OF TEST EXCEEDED THREE HOURS, THERE IS A HIGH POSSIBILITY THAT THE DISCREPANCIES ARE DUE TO CHANGES IN THE STATUS OF THE PATIENT. THREE HOURS IS CONSIDERED A REASONABLE LENGTH OF TIME BETWEEN TWO READINGS IN ORDER FOR COMPARISON TO BE VALID. PREVIOUS INVESTIGATION ON STRIP LOT #222167 FROM A PREVIOUS CASE MET ACCURACY CRITERIA. NO FURTHER INVESTIGATION WILL BE PURSUED. INVESTIGATION RESULT: THE ALLOWABLE DIFFERENCE BETWEEN THE INRATIO INR'S AND SYSMEX INR'S ARE CALCULATED. AT LEAST TWO OUT THREE REPLICATES FOR BOTH SAMPLES FOR LOT #222167 ARE WITHIN THE ALLOWABLE BIAS. NO DISCREPANT RESULTS WERE ESTABLISHED ON RETURNED AND IN-HOUSE METER. THESE MEET THE ABOVE CRITERIA, AND THEN NO FURTHER ACTION IS REQUIRED. CONCLUSION: BASED ON CUSTOMER'S REPORTED TEST DATA, DATA ANALYSIS OF MEAN CALCULATION REVEALED ACCURACY CRITERIA WAS NOT MET. NO PRODUCT IS EXPECTED TO BE RETURNED, RETAIN STRIP TEST DATA FROM A PREVIOUS CASE INDICATED PRODUCT DEFICIENCY NOT ESTABLISHED ON RETAIN STRIP LOT# 222167. PATIENT CONDITION AND MEDICATIONS MAY AFFECT INRATIO INR TEST. THERE IS INSUFFICIENT INFORMATION TO DETERMINE THE ROOT CAUSE OF CUSTOMER'S TEST RESULT DISCREPANCY. AS OF 03/30/2010, 5 DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #222167 YIELDING A COMPLAINT RATE OF 0.001%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ON GOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B) (6) 2010, INRATIO: 2.8, LAB: 1.2. DATE: (B) (6) 2010, LAB: 1.5. DATE: (B) (6) 2010, LAB: 1.6. DATE: (B) (6) 2010, LAB: 2.2. DATE: (B) (6) 2009, LAB: 2.6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | BIOSITE INCORPORATED | 100139 | 222167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |