FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 16540044 · Received March 14, 2023

Report

Report Number
3013756811-2023-34412
Event Type
Malfunction
Date Received
March 14, 2023
Date of Event
January 8, 2023
Report Date
March 14, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00853052007301
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. CUSTOMER LOADED A NEW CARTRIDGE TO RESOLVE THE ISSUE. CUSTOMERS BLOOD GLUCOSE LEVEL WAS 150-175 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
935079 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00853052007301

Patients

Seq Age Sex Outcome Treatment
1 78 YR Male