FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY

MDR report key: 16539309 · Received March 14, 2023

Report

Report Number
3013756811-2023-37471
Event Type
Malfunction
Date Received
March 14, 2023
Date of Event
February 19, 2023
Report Date
May 17, 2023
Manufacturer
TANDEM DIABETES CARE
Product Code
QFG
UDI-DI
00850006613205
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE FAILURE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, THE ALLEGED BATTERY ISSUE WAS VERIFIED, BUT THE MINIMUM FILL ISSUE COULD NOT BE VERIFIED.

Additional Manufacturer Narrative · 0

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 250 UNITS OF INSULIN DURING THE LOAD SEQUENCE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 180-182 MG/DL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
914790 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1002717 00850006613205

Patients

Seq Age Sex Outcome Treatment
1 21 YR Female INSULIN: HUMALOG