FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 16539309
·
Received March 14, 2023
Report
- Report Number
- 3013756811-2023-37471
- Event Type
- Malfunction
- Date Received
- March 14, 2023
- Date of Event
- February 19, 2023
- Report Date
- May 17, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613205
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE FAILURE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, THE ALLEGED BATTERY ISSUE WAS VERIFIED, BUT THE MINIMUM FILL ISSUE COULD NOT BE VERIFIED.
Additional Manufacturer Narrative · 0
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 250 UNITS OF INSULIN DURING THE LOAD SEQUENCE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 180-182 MG/DL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 914790 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Female | INSULIN: HUMALOG |