FDA Adverse Event Malfunction Summary report: N

BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK

MDR report key: 16538281 · Received March 14, 2023

Report

Report Number
1119779-2023-00279
Event Type
Malfunction
Date Received
March 14, 2023
Date of Event
February 22, 2023
Report Date
March 15, 2023
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: MATERIAL #: 251181. LOT/BATCH #: 2325441. BD RECEIVED SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE INDICATED FAILURE MODE FOR FOREIGN MATTER WITH THE INCIDENT LOT WAS NOT OBSERVED. THE RETENTION SAMPLES COULD NOT BE REVIEWED AS THE BATCH WAS ALREADY EXPIRED. BASED ON A REVIEW OF THE BATCH HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS NOT CONFIRMED FOR THE INDICATED FAILURE MODE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK A FUZZY APPEARANCE WAS FOUND IN MEDIA BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THIS IS A REPORT ABOUT FOREIGN MATTER. THE CUSTOMER FOUND MATTER WITH A FUZZY APPEARANCE IN THE MEDIA BEFORE USAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK A FUZZY APPEARANCE WAS FOUND IN MEDIA BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THIS IS A REPORT ABOUT FOREIGN MATTER. THE CUSTOMER FOUND MATTER WITH A FUZZY APPEARANCE IN THE MEDIA BEFORE USAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
914717 BD BBL¿ SALMONELLA SHIGELLA AGAR PLATE 20PK CULTURE MEDIA, SELECTIVE AND DIFFERENTIAL JSI BECTON, DICKINSON & CO. (SPARKS) 2325441

Patients

Seq Age Sex Outcome Treatment
1 Unknown