FDA Adverse Event Malfunction Summary report: N

0998-00-0600-2054U

MDR report key: 1653394 · Received March 30, 2010

Report

Report Number
2221819-2010-00007
Event Type
Malfunction
Date Received
March 30, 2010
Date of Event
January 25, 2010
Report Date
January 26, 2010
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6) SENT PHOTOGRAPHS TO MINDRAY THAT SHOW THAT THE PLASTIC HINGE BARREL BROKE ON THE MOUNT IN QUESTION, WHICH IS FIVE YRS OLD. THIS MOUNTING ASSEMBLY IS CURRENTLY MFG BY (B)(4) FOR MINDRAY DS USA, INC (FORMERLY DATASCOPE PATIENT MONITORING AND PRIOR TO THAT, DATASCOPE CORP.). THE MOUNTING ASSEMBLY THAT FAILED AT (B)(6) WAS MFG FOR DATASCOPE CORP. BY (B)(4). THE CHANGE IN SUPPLIER (B)(4) WAS MADE FOR BUSINESS RELATED REASONS. DURING THE DESIGN TRANSFER PROCESS, AN IMPROVEMENT WAS MADE TO THE DESIGN OF THE LATCH. THE PLASTIC THAT HOLDS THE PIN TO THE MOUNT WAS STRENGTHENED. MOUNTS WITH THE IMPROVED LATCH DESIGN ARE ON ORDER, AND THE FIRST SHIPMENT IS EXPECTED TO BE RECEIVED SHORTLY. (B)(4).

Description of Event or Problem · 1

WE RECEIVED AN INQUIRY FROM (B)(6) RELATED TO THE FOLLOWING EVENT: "BASE PLATE CLAMP ON ROLLING STAND THAT SECURES MONITOR TO IT HAS BROKEN OFF. MONITOR THEREFORE PROJECTED OFF STAND AND SMASHED ONTO FLOOR NARROWLY MISSING STAFF AND PTS. RESULTS BEING THAT THE DISPLAY AND CASING DAMAGED OTHER INTERNAL ISSUES YET KNOWN. UNIT REMOVED FROM WARD AREA. BASE PLATE RETAINED WITHIN EBME DEPT AND THE REPAIR PROCESS STARTED FOR THE MONITOR." NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 0998-00-0600-2054U TRIO MHX MINDRAY DS USA, INC. TRIO NA

Patients

Seq Age Sex Outcome Treatment
1 UNK