FDA Adverse Event
Malfunction
Summary report: N
ROADRUNNER
MDR report key: 16532
·
Received September 30, 1994
Report
- Report Number
- MW1003557
- Event Type
- Malfunction
- Date Received
- September 30, 1994
- Date of Event
- June 1, 1994
- Report Date
- September 23, 1994
- Manufacturer
- AIR-SHIELDS VICKERS, INC.
- Product Code
- FPL
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
DURING TRANSPORT, DC POWER WAS LOST. OF THE FOUR CLIPS ON THE TWO BATTERIES, THREE CLIPS WRE FOUND TO BE DISCONNECTED. RPTR FOUND THAT THE CLIPS' OPENING HAD WIDENED AND LOOSENED. RPTR CHECKED TWO OTHER INCUBATORS AND FOUND THAT ALL THE CLIPS ON ONE WERE VERY LOOSE AND OFFERED NO RESISTANCE TO REMOVAL. RPTR BELIEVES THAT THE CLIPS USED MAY BE INADEQUATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ROADRUNNER | TRANSPORT INCUBATOR | FPL | AIR-SHIELDS VICKERS, INC. | TI 100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |