FDA Adverse Event Malfunction Summary report: N

ROADRUNNER

MDR report key: 16532 · Received September 30, 1994

Report

Report Number
MW1003557
Event Type
Malfunction
Date Received
September 30, 1994
Date of Event
June 1, 1994
Report Date
September 23, 1994
Manufacturer
AIR-SHIELDS VICKERS, INC.
Product Code
FPL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

DURING TRANSPORT, DC POWER WAS LOST. OF THE FOUR CLIPS ON THE TWO BATTERIES, THREE CLIPS WRE FOUND TO BE DISCONNECTED. RPTR FOUND THAT THE CLIPS' OPENING HAD WIDENED AND LOOSENED. RPTR CHECKED TWO OTHER INCUBATORS AND FOUND THAT ALL THE CLIPS ON ONE WERE VERY LOOSE AND OFFERED NO RESISTANCE TO REMOVAL. RPTR BELIEVES THAT THE CLIPS USED MAY BE INADEQUATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROADRUNNER TRANSPORT INCUBATOR FPL AIR-SHIELDS VICKERS, INC. TI 100

Patients

Seq Age Sex Outcome Treatment
1 *