FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 16531513 · Received March 13, 2023

Report

Report Number
3014704491-2023-00048
Event Type
Malfunction
Date Received
March 13, 2023
Date of Event
January 10, 2023
Report Date
March 1, 2023
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED, A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2172398. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED AT THE CONNECTION SITE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "ON (B)(6) 2023, THE PATIENT WAS TREATED WITH A CLOSED VENOUS INDWELLING NEEDLE, AND LEAKAGE WAS FOUND AT THE JOINT AT THE LOWER END OF THE HEPARIN CAP WHEN EXHAUSTING. IF FOUND, REPLACE THE INDWELLING NEEDLE IMMEDIATELY AND REPORT TO THE EQUIPMENT DEPARTMENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1649871 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 2172398

Patients