FDA Adverse Event Injury Summary report: N

FAST-FIX 360 CURVED NDL DELIVERY SYS

MDR report key: 16530431 · Received March 13, 2023

Report

Report Number
1219602-2023-00366
Event Type
Injury
Date Received
March 13, 2023
Date of Event
February 21, 2023
Report Date
October 10, 2023
Manufacturer
SMITH & NEPHEW, INC.
Product Code
GAT
UDI-DI
00885554023077
PMA / PMN Number
K092508
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: (B)(4). THE REPORTED DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION REVEALED THAT IT IS NOT IN ITS ORIGINAL PACKAGING. THE INSERTION DEVICE WAS RETURNED IN THE PRET1 SETTING WITH NO SUTURE OR IMPLANTS RETURNED. A REVIEW OF THE CUSTOMER SUPPLIED IMAGE FINDS THE COMPLAINT DEVICE ON A SMITH+NEPHEW PRODUCT BOX WITH A LENGTH OF SUTURE BUT NO ANCHORS VISIBLE. A FUNCTIONAL EVALUATION WAS PERFORMED ON THE RETURNED DEVICE AND FOUND IT CYCLES AS DESIGNED. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATIONS UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. THE INSTRUCTIONS FOR USE WERE REVIEWED AND FOUND TO INCLUDE CONDITIONS OF OFF LABEL USE AND TECHNIQUE SPECIFICS, AS WELL AS PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THE ROOT CAUSE COULD NOT BE DETERMINED SINCE THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED DURING THE INVESTIGATION. FACTORS THAT CAN CONTRIBUTE TO THE REPORTED EVENT INCLUDE AN APPLICATION OF EXCESSIVE FORCE OR CONTACT WITH ANOTHER SOURCE. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.

Additional Manufacturer Narrative · 0

INTERNAL COMPLAINT REFERENCE: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A MENISCUS REPAIR SURGERY, T2 OF THE FAST-FIX 360 COULD NOT BE DEPLOYED. T1 WAS REMOVED FROM THE PATIENT BY TWEEZERS. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
593194 FAST-FIX 360 CURVED NDL DELIVERY SYS SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE GAT SMITH & NEPHEW, INC. 72202468 2080912 00885554023077

Patients

Seq Age Sex Outcome Treatment
1 35 YR Male Required Intervention