FDA Adverse Event Malfunction Summary report: N

ACUJECT VARIABLE INJECTION NEEDLE

MDR report key: 1652984 · Received March 30, 2010

Report

Report Number
1037905-2010-00144
Event Type
Malfunction
Date Received
March 30, 2010
Date of Event
March 1, 2010
Report Date
March 1, 2010
Manufacturer
COOK ENDOSCOPY
Product Code
FBK
PMA / PMN Number
K941305
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVALUATION. WE COULD NOT CONDUCT A REVIEW OF THE DEVICE HISTORY RECORD OR CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE THE LOT NUMBER WAS NOT PROVIDED. AFTER A REVIEW OF THE ACCOUNT ORDERING AND SHIPPING HISTORY, THE LOT NUMBER COULD NOT BE DETERMINED. A REVIEW OF THE TWELVE MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS RARE. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. NEEDLE EXTENSION DIFFICULTY CAN OCCUR IF NEEDLE EXTENSION IS ATTEMPTED WITH THE CATHETER IN A COILED OR CURVED POSITION. THE INSTRUCTIONS FOR USE DIRECT THE USER TO UNCOIL THE CATHETER AND STRAIGHTEN COMPLETELY. THE INSTRUCTIONS FOR USE ALSO CAUTION THE USER THAT EXTENDING THE NEEDLE WHILE THE CATHETER IS COILED MAY RESULT IN DAMAGE TO THE PERFORMANCE CHARACTERISTICS OF THE DEVICE. PRIOR TO DISTRIBUTION, ALL ACUJECT VARIABLE INJECTION NEEDLES ARE SUBJECTED TO A FUNCTIONAL TEST TO ENSURE APPROPRIATE NEEDLE EXTENSION. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS RARE. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK ENDOSCOPY ACUJECT VARIABLE INJECTION NEEDLE. DURING THE PROCEDURE, THE NEEDLE WOULD NOT EXTEND FROM THE OUTER CATHETER. ANOTHER MFR'S DEVICE WAS USED TO COMPLETE THE PROCEDURE. A FOREIGN OBJECT DID NOT HAVE TO BE RETRIEVED FROM THE PT. NO ADDITIONAL MEDICAL PROCEDURES WERE REQUIRED DUE TO THIS OCCURRENCE. THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUJECT VARIABLE INJECTION NEEDLE FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY FBK COOK ENDOSCOPY UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK OLYMPUS GIF-160 ENDOSCOPE