FDA Adverse Event Injury Summary report: N

COCR BALL HEAD 12/14 Ø 36 SIZE M 0

MDR report key: 16529554 · Received March 13, 2023

Report

Report Number
3005180920-2023-00062
Event Type
Injury
Date Received
March 13, 2023
Date of Event
February 16, 2023
Report Date
March 13, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805219
PMA / PMN Number
K080885
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 23 FEBRUARY 2023: LOT 161630C: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-MAY-2021. EXPIRATION DATE: 2026-05-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. LOT 161630: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-MAY-2016. EXPIRATION DATE: 2021-04-13. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICE INVOLVED. BATCH REVIEW PERFORMED ON 23 FEBRUARY 2023: LINER: MPACT 01.32.3644HCAT HOODED PE HC LINER Ø36/E (K132879) LOT 2202312: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 03-MARCH-2022. EXPIRATION DATE: 2027-02-13. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

REVISION SURGERY PERFORMED AT ABOUT 8 MONTHS AFTER PRIMARY DUE TO JOINT LUXATION (HEAD FROM LINER). THE SURGEON REVISED SUCCESSFULLY THE HEAD AND LINER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
576542 COCR BALL HEAD 12/14 Ø 36 SIZE M 0 HIP COCR BALL HEAD JDI MEDACTA INTERNATIONAL SA 01.25.031 161630C 07630030805219

Patients

Seq Age Sex Outcome Treatment
1 81 YR Male Required Intervention