INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE
Report
- Report Number
- 1037905-2023-00122
- Event Type
- Injury
- Date Received
- March 13, 2023
- Date of Event
- February 13, 2023
- Report Date
- March 13, 2023
- Manufacturer
- WILSON-COOK MEDICAL INC
- Product Code
- PKL
- UDI-DI
- 00827002580107
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PMA/510(K): K212323. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE STATES, "NOTE: IF SEPARATION OF CLIP IS NOT IMMEDIATE, GENTLY MOVE CATHETER BACK AND FORTH OR USE OTHER ENDOSCOPIC MANEUVERS TO SEPARATE CATHETER FROM CLIP." PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING A COLONOSCOPY, THE PHYSICIAN USED THREE COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. IT WAS REPORTED THAT THE CLIPS WERE ATTACHED TO THE SITE AND WOULD NOT RELEASE FROM THE DRIVE WIRE. THE CLIP HAD TO BE PULLED OFF THE SITE WHILE THE CLIP WAS STILL CLOSED AND ATTACHED. THE COOK REPRESENTATIVE IN CONTACT WITH THE CLINICAL SITE STATED THAT THE FINAL CLIP MAY HAVE TREATED THE PERFORATION WHILE ALSO CLIPPING THE INITIAL LESION. THE PATIENT WAS KEPT AT THE HOSPITAL FOR OBSERVATION. NO OTHER ADVERSE EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449200 | INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE | PKL, LIGATOR, HEMORRHOIDAL | PKL | WILSON-COOK MEDICAL INC | G58010 | W4664675 | 00827002580107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | OLYMPUS PCF-190 COLONOSCOPE. |