FDA Adverse Event Injury Summary report: N

INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE

MDR report key: 16529085 · Received March 13, 2023

Report

Report Number
1037905-2023-00122
Event Type
Injury
Date Received
March 13, 2023
Date of Event
February 13, 2023
Report Date
March 13, 2023
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
PKL
UDI-DI
00827002580107
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K): K212323. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE INSTRUCTIONS FOR USE STATES, "NOTE: IF SEPARATION OF CLIP IS NOT IMMEDIATE, GENTLY MOVE CATHETER BACK AND FORTH OR USE OTHER ENDOSCOPIC MANEUVERS TO SEPARATE CATHETER FROM CLIP." PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING A COLONOSCOPY, THE PHYSICIAN USED THREE COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. IT WAS REPORTED THAT THE CLIPS WERE ATTACHED TO THE SITE AND WOULD NOT RELEASE FROM THE DRIVE WIRE. THE CLIP HAD TO BE PULLED OFF THE SITE WHILE THE CLIP WAS STILL CLOSED AND ATTACHED. THE COOK REPRESENTATIVE IN CONTACT WITH THE CLINICAL SITE STATED THAT THE FINAL CLIP MAY HAVE TREATED THE PERFORATION WHILE ALSO CLIPPING THE INITIAL LESION. THE PATIENT WAS KEPT AT THE HOSPITAL FOR OBSERVATION. NO OTHER ADVERSE EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449200 INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE PKL, LIGATOR, HEMORRHOIDAL PKL WILSON-COOK MEDICAL INC G58010 W4664675 00827002580107

Patients

Seq Age Sex Outcome Treatment
1 Unknown OLYMPUS PCF-190 COLONOSCOPE.